Teledyne 3190 manual Gas Connections, Installation Checklist

Page 25

Trace Oxygen Analyzer

Installation 3

 

 

 

 

 

 

 

 

 

is keyed to fit only one way into the receptacle. Do not force it in. The other end of the cable is made of four separate wires. These should be connected to the terminal strip on the analysis unit as follows:

Red:

#1

 

Black:

#2

 

Green:

#3}

The green and white connectors can be

White:

#4

interchanged, but be consistent.

Refer to the Final Assembly, Dwg. C-64641.

3.4Gas Connections

Gas connection instructions depend on the specific External Probe used and any special requirements of the process being monitored.

The standard Model 3190 External Probe has inlet and outlet fixtures only. Calibration gasses must be tee'd into the sample inlet through appropri- ate valves. ¼ inch tube fittings are used. For metric installations, ¼ inch to 6 mm adapters are supplied.

In general, sample flow and pressure must not create significant backpressure past the sensor. For the standard probe, 2 scfh is the nominal recommended flowrate.

The pressure required will depend on the sampling system. When venting into a constant pressure, such as the atmosphere, controlling input pressure is simple. If you are venting into a system of varying pressure, then some form of pressure regulation is required.

3.5Installation Checklist

Before connecting the instrument to the power source and turning it on, make sure you have:

Correctly installed the Sample and Exhaust gas lines

Opened the isolation valves

Checked for leaks

Set the sample pressure to 5–10 psig, nominal

Once the above checks have been made, you can connect to the power source. The instrument is now ready for operation.

Teledyne Analytical Instruments

3-7

Image 25
Contents Teledyne Analytical Instruments Model 3190 SeriesCopyright 1999 Teledyne Analytical Instruments Specific Model Information Contents Appendix MaintenanceTeledyne Analytical Instruments Introduction Main Features of the AnalyzerOverview Trace Oxygen Analyzer IntroductionIntroduction Model Front Panel DescriptionUp Arrow Rear Panel DescriptionRear Panel AC and DC versions Analog OutputsRS-232 Port Introduction Model Teledyne Analytical Instruments Introduction Operational TheoryMicro-Fuel Cell Sensor Trace Oxygen Analyzer Operational TheoryAnatomy of a Micro-Fuel Cell Operational Theory Model+ 2H 2O + 4e → 4OH Electrochemical ReactionsEffect of Pressure Calibration CharacteristicsElectronics Signal ProcessingGeneral Block Diagram of the Signal Processing Electronics Trace Oxygen Analyzer Installation InstallationUnpacking the Analyzer Control Unit Installation Installation ModelExternal Probe Installation Location and MountingElectrical Connections Installing the Micro-Fuel Cell / Cell Block OrientationRear Panel Electrical Connectors for AC and DC Units Trace Oxygen Analyzer Installation LO Alarm HI AlarmParameter Setting Sensor FailGas Connections Installation ChecklistInstallation Model Teledyne Analytical Instruments Operation Using the Function and Data EntryTrace Oxygen Analyzer Operation Buttons Setting the Analysis RangesOperation Model Settle Mode Setting the Alarm SetpointsHI Alarm Trace Oxygen Analyzer Operation HI RangeSensor Fail Alarm LO AlarmSelecting a Fixed Range or Autoranging CalibrationDisplaying Percent & PPM on the LED Display ProcedureSetL mode in the LED display Trace Oxygen Analyzer Maintenance MaintenanceAC Powered Units Replacing the FuseMaintenance Model Sensor Installation or ReplacementDC Powered Units When to Replace a SensorRemoving the Micro-Fuel Cell Disconnect the Power Source at the Control UnitExploded View of MFC and Cell Block Installing a Micro-Fuel CellCell Warranty Conditions Disconnect the Power Source from the Control UnitMaintenance Model Teledyne Analytical Instruments Trace Oxygen Analyzer Appendix Appendix SpecificationsAppendix Model Spare Parts ListMiscellaneous Drawing ListAppendix Model Teledyne Analytical Instruments