3.6.1.2. M501-TRS Alarm Output Connections

M102E/M501 TRS

GETTING STARTED

(Addendum to M101E Manual - P/N 04740 Rev A)

TO change these settings, see Sections 6.9.4 and 6.7 of the M101E Manual - P/N 04740 Rev A respectively.

3.6.1.2. M501-TRS Alarm Output Connections

The rear panel of the M501-TRS includes a terminal strip by which connections can be made to the converters internal temperature alarm for more information on this alarm see Section 6.5).

Connect the input leads to your alarm-sensing device (e.g. datalogger) to the center two pins of the alarm output connector (see Figure 3-5).

Make sure the load does not exceed the rated capacity of the relay.

3.6.2. Pneumatic Connections:

This section supercedes the information contained in Section 3.1.2 of the M101E Manual - P/N 04740 Rev A.

CAUTION

To prevent dust from getting into the analyzer, it was shipped with small plugs inserted into each of the pneumatic fittings on the rear panel. Make sure that all dust plugs are removed before attaching exhaust and supply gas lines.

Sample and calibration gases should only come into contact with PTFE (Teflon) or glass materials. They should not come in contact with FEP or stainless steel materials.

Figures 3-7 and 3-8 show the most common configurations for gas supply and exhaust lines to the Model 102E Analyzer. Figures 3-9, 3-10 & 3-11 show the connections for units with valve options installed.

Please refer to Figures 3-1 & 3.3 for the location of pneumatic connections at the rear panel of the M102E and the M501-TRS.

Table 3-3: Inlet / Outlet Connector Nomenclature

 

 

 

M102E PNEMATIC CONNECTERS

 

REAR PANEL LABEL

 

FUNCTION

 

SAMPLE

 

Connects the sample gas to the analyzer. When operating the analyzer without

 

 

zero/span option, this is also the inlet for any calibration gases.

 

 

 

 

 

 

 

 

EXHAUST

 

Exhausts the gas sampled by the analyzer. Connect to an outside area away

 

 

from people.

 

 

 

 

 

 

 

 

SPAN

 

On units with zero/span/shutoff valve options installed, connect a gas line to

 

 

the source of calibrated span gas here.

 

 

 

 

 

 

 

 

ZERO AIR

 

On Units with zero/span valve or IZS option installed, this port connects the

 

 

zero air gas or the zero air cartridge to the analyzer.

 

 

 

 

 

 

 

 

TO CONVERTER

 

Sample gas leaves the M102E to be conditioned by the M501-TRS via this port.

 

 

 

 

 

FROM CONVERTER

 

Sample gas returns to the M102E after being conditioned by the M501-TRS via

 

 

this port.

 

 

 

 

 

 

 

 

 

 

M501-TRS PNEMATIC CONNECTERS

 

REAR PANEL LABEL

 

FUNCTION

 

 

 

 

05514 Rev A1

11

Image 17
Contents MODEL 501 TRS THERMAL CONVERTER MODEL 102E TOTAL REDUCED SULFUR ANALYZERADDENDUM TO MODEL 101E OPERATORS MANUAL withPage M102E/M501 TRS TABLE OF CONTENTS05514 Rev A1 Addendum to M101E Manual - P/N 04740 Rev A8. INSTRUMENT MAINTENANCE LIST OF APPENDICES10. TROUBLESHOOTING & REPAIR 9. THEORY OF OPERATIONPREFACE LIST OF FIGURESLIST OF TABLES Page corner of the display any time the instrument is in SETUP mode 1. PREFACE1.1. Reference Numbering convention User Notes2.1. Specifications 2. SPECIFICATIONS, APPROVALS AND WARRANTY2.1.1. M501-TRS Specifications 2.2. EPA Equivalency Designation2.3. CE Mark Compliance 3.1. Unpacking the M102E 3. GETTING STARTED3.2. Unpacking the M501-TRS Never disconnect electronic circuit boards, wiring harnesses orthe maximum operating temperature specification for the M102E 40C 3.2.1. M501-TRS Ventilation Clearancethat the rack/enclosure itself is adequately ventilated Figure 3-2 M501-TRS Internal Layout 3.3. Internal LayoutsFigure 3-1 M102E Internal Layout Table 3-1 TRS - SO2 Switching Valve Operating Modes 3.4. Internal Pneumatic Flow of the M102E & the M501-TRSM102E INSTRUMENT CHASSIS M501Alarm Output 3.5. Rear Panel Layout for the M102E & M501-TRSFigure 3-4 M102E Rear Panel Layout 5 Amp3.6.1. Electrical Connections 3.6. Initial Setup3.6.1.1. M102E Analog Output Connections Figure 3-6 Analog Output Connector3.6.2. Pneumatic Connections 3.6.1.2. M501-TRS Alarm Output ConnectionsM102E PNEMATIC CONNECTERS M501-TRS PNEMATIC CONNECTERS102E MODELSPAN GAS The exhaust from the instrument needs to be vented outside theZERO AIR 3.6.2.1. Connections with Internal Valve Options InstalledVENT if input is pressurized Gas Dilution CalibratorVENT Valve Option Option3.7.1. Startup / Warm Up of the M102E 3.7. Initial Operation3.7.2. Functional Check of the M102E Possible Warning Messages at Start-Up3.7.3. Startup / Warm Up of the M501-TRS High Alarm LED Low Alarm LEDProcess Variable Display area3.8. Initial Calibration Page 4.1. Rack Mount Kits Options 20a, 20b, 21, 22 4. OPTIONAL HARDWARE AND SOFTWARE4.2. Calibration Valves Options OPTIONAL HARDWARE AND SOFTWAREZero/Span Valve Operating States Table 4-2 IZS Valve Operating States M102E INSTRUMENT CHASSIS4.3.2. Addendum on CD Part number 4.3. Additional Manuals4.3.1. Printed Manuals P/N 5. M102E OPERATING INSTRUCTIONS 5.1.1. M102E Analog Output SignalsFigure 5-1 Analog Output Connector Key M102E OPERATING INSTRUCTIONS5.1.2. Setting the M102E Gas Measurement Mode 5.2. SETUP - DIAG Using the Diagnostics Functions5.2.1. M102E Analog I/O Configuration Table 5-1 M102E gas Measurement Modes5.2.2. M102E Test Channel Output 5.3. SETUP - COMM Setting Up the M102E’s Communication Ports5.3.1. M102E ID Code 5.3.2. M102E Ethernet Host Name5.4.1. Control Inputs 5.4. Remote Operation of the AnalyzerCONTROL IN ZERO5 VDC Power Supply 5.4.2. Using the M102E with a Hessen Protocol Network5.4.2.1. M102E Hessen Protocol Gas ID List SPANTable 5-6 Default Hessen Status Bit Assignments 5.4.2.2. Setting Hessen Protocol Status Flags6.1. Basic M501-TRS Controls 6. M501-TRS OPERATING INSTRUCTIONSDO NOT OPERATE WITHOUT THE COVER OF THE M501TS CONVERTER INSTALLED Teledyne Instruments customer serviceTable 6-1 M501-TRS Temperature Controls and Definitions 6.2. To Display The Current TemperatureDO NOT SET THE TEMPERATURE HIGHER THAN 1050OC 6.3. To Manually Adjust the Converter Oven Temperaturehas reached a stable, constant temperature 6.4. Autotune the Temperature Controller6.4.1. Initiating the Autotune Process necessary to repeat the autotune procedure 6.5. M501TRS Alarm Relay Adjustment6.4.2. Aborting the Autotune Process Page USER NOTES 7. CALIBRATION PROCEDURES7.1. M102E Calibration 7.2. M501-TRS CalibrationPage INSTRUMENT MAINTENANCE 8. INSTRUMENT MAINTENANCE8.1.1. Maintaining the SO2 Scrubber 8.1. Additional and Updated Maintenance Procedures8.1.1.1. Predicting When the SO2 Scrubber Should Be Replaced 8.1.1.3. Changing the SO2 Scrubber Material 8.1.1.2. Checking the Function of the SO2 ScrubberPage 9.1. Measurement Principle 9. THEORY OF OPERATION9.1.1. TRS Conversion 9.1.2. SO2 Ultraviolet FluorescenceSO2 * ⎯⎯→ SO2 + hv330nm SO2 + hv214nm ⎯⎯Ia⎯→ SO2Ia = I0 1 − exp− axSO2 Figure 9-1 UV Absorption in the M102E Reaction CellEquation9-6 F = kSO2kSO2 * ⎯⎯F⎯→ SO2 + hv330nm 9.3. Pneumatic Operation 9.2. The UV Light Path9.2.1. UV Lamp Shutter & PMT Offset 9.3.1. Sample gas Flow9.4. Electronic Operation PMT HOUSING9.4.1. Sensor Module SAMPLE CHAMBERUV Lamp9.4.1.2. Sample Chamber Heating Circuit 9.4.1.1. Sample ChamberSample Air Housing Outlet O-Ring O-Ring Seal Seal Sample Air9.4.2. M501-TRS electronics 9.4.2.1. Thermal SwitchHEATER P-I-D CONTROLLER9.4.2.2. Temperature Alarms and Alarm Output 10.1.1. Fault Diagnosis with Warning Messages 10. TROUBLESHOOTING & REPAIR10.1.1.2. M501-TRS Error Codes 10.1.1.1. M102E Warning Messagesfunctioning 10.1.2. Fault Diagnosis with Test FunctionsTable 10-2 Test Functions - Possible Causes for Out-Of-Range Values 10.3. Other Performance Problems 10.2. M501-TRS Trouble shooting10.2.1. TRS Converter Not Heating 10.3.1. Excessive noise10.4.2. Checking the Efficiency of the M501-TRS TRS Æ SO2 Converter 10.4. Subsystem Checkout10.4.1. Checking the Efficiency of the M501-TRS SO2 Scrubber 10.5.1. UV Lamp Adjustment and/or Replacement 10.5. Additional Repair ProceduresALWAYS wear UV-Protective, Safety Glasses when working with the UV 10.5.1.1. Adjusting the UV Lamp Peaking the LampLamp Assembly Always grasp the main body of the lampFigure 10-1 Shutter Assembly - Exploded View 10.5.1.2. Replacing the UV LampFigure 10-2 Disassembling the Shutter Assembly 10.5.2. Replacing the UV filter/lensFigure 10-3 PMT Assembly - Exploded View 10.5.3. Replacing the PMT, HVPS or TEC3. Remove the reaction cell assembly 10.5.4. M102E PMT Hardware Calibration FACTORY CAL Figure 10-4 Pre-Amplifier Board Layout 6. Locate the Preamp board see FigureIf a reporting range other than 500 ppb is used in this procedure 10.5.5. Replacing the TRS Converter Heating TubeAdjust the PMT until NORM PMT equals 1280 mV ± 10 mV NORM PMT value of twice the ppb value of the span gas EXAMPLEThe ceramic bobbins at each end of the heater assembly are fragile 10.6. Manually Programming the M501-TRS Temperature ControllerBe Careful element off115V/60Hz with a set value of 1000C 10.6.1. Temperature Controller Primary Menu ParametersTable 10-3 - Temperature Controller - Primary Parameter Settings Table 10-4 - Temperature Controller - Primary Parameter Settings 10.7. Technical AssistanceUser Notes APPENDIX A-2 Model 102E Setup Variables Available Via Serial I/O APPENDIX A - Version Specific Software DocumentationAPPENDIX A-4 Model 102E Signal I/O Definitions M102E/M501 TRS APPENDIX A - Version Specific Software DocumentationENTER SETUP PASS SETUPAPPENDIX A-1 M102E Software Menu Trees, Revision A.2 APPENDIX A-1 M102E Software Menu Trees, Revision A.2 SAMPLETEST SAMPLEMSG 1,2 CLR 1,3TR2, 112, REPORTED ACAL1SO2, 111, REPORTED DIAG SAMPLE ENTER SETUP PASSFigure A-4 Secondary Setup Menu DIAG Deleted Setup Variables for M102E Software Revision A.2 APPENDIX A-2 Setup Variables For Serial I/O, Revision A.2Table A-2 MEASUREMODEAPPENDIX A-3 Warnings and Test Functions, Revision A.2 setting of STABILFREQ and STABILSAMPLESTable A-4 Warning Messages deleted from M102 Software Revision A.2 Table A-6 Test Functions Deleted from M102 Software Revision A.2Rear board primary MUX analog inputs APPENDIX A-4 M102E Signal I/O Definitions, Revision A.2APPENDIX A-4 M102E Signal I/O Definitions, Revision A.2M102E/M501 TRS Trigger Event APPENDIX A-5 M102E iDAS Functions, Revision A.2CTEMPW FunctionAPPENDIX A-5 M102E iDAS Functions, Revision A.2 A-10APPENDIX B - M102E Spare Parts List APPENDIX B - M102E Spare Parts ListTable B-1 M102E Spare Parts List M101E Manual - P/N 04740120 Rev AM102E/M501 TRS M102E/M501 TRS Warranty/Repair Questionnaire Model 102ETELEDYNE INSTRUMENTSM102E/M501 TRS APPENDIX D - ELECTRONIC SCHEMATICS APPENDIX D - ELECTRONIC SCHEMATICSTable D-1 List of Included Electronic Schematics Addendum to M101E Manual - P/N 04740 Rev AUser Notes APPENDIX D - ELECTRONIC SCHEMATICSM102E/M501 TRS 05518 Rev A