Figure 10-1: Shutter Assembly - Exploded View

M102E/M501 TRS

TROUBLESHOOTING & REPAIR

(Addendum to M101E Manual - P/N 04740 Rev A)

Reaction

10.5.1.2. Replacing the UV Lamp Cell

DO NOT

UV Filter Retainer

& Lens Housing

Shutter Housing

use Lamp Cap to

 

adjust Lamp

 

position

UV Lamp Power

Supply Wires

 

 

Adjust Lamp

 

 

Position by

 

 

grasping lamp

 

 

body ONLY

 

 

 

 

 

Thumb

Shutter Assy

 

Screw

 

 

 

UV Lamp Bracket

 

Mounting Screws

Figure 10-1: Shutter Assembly - Exploded View

10.5.1.2. Replacing the UV Lamp

1.Turn off the analyzer.

2.Disconnect the UV lamp from its power supply.

You can find the power supply connector by following the two, white UV Lamp power supply wires from the lamp to the power supply.

3.Loosen, but do not remove the two UV lamp bracket screws, and the large brass thumbscrew located on the shutter housing (see Figure 10-1) so that the lamp can be moved.

NOTE:

DO NOT grasp the UV lamp by its cap when changing its position (see Figure 10-1).

Always grasp the main body of the lamp.

4.Remove the UV Lamp by pulling it straight up.

5.Insert the new UV lamp into the bracket.

6.Tighten the two UV lamp bracket screws, but leave the brass thumb screw un-tightened.

7.Connect the new UV lamp to the power supply.

8.Turn the instrument on and perform the UV adjustment procedure as defined in section 10.5.1.1 of this addendum

05514 Rev A1

57

Image 63
Contents with MODEL 102E TOTAL REDUCED SULFUR ANALYZERMODEL 501 TRS THERMAL CONVERTER ADDENDUM TO MODEL 101E OPERATORS MANUALPage Addendum to M101E Manual - P/N 04740 Rev A TABLE OF CONTENTSM102E/M501 TRS 05514 Rev A19. THEORY OF OPERATION LIST OF APPENDICES8. INSTRUMENT MAINTENANCE 10. TROUBLESHOOTING & REPAIRLIST OF FIGURES LIST OF TABLESPREFACE Page corner of the display any time the instrument is in SETUP mode 1. PREFACE1.1. Reference Numbering convention User Notes2.2. EPA Equivalency Designation 2. SPECIFICATIONS, APPROVALS AND WARRANTY2.1. Specifications 2.1.1. M501-TRS Specifications2.3. CE Mark Compliance Never disconnect electronic circuit boards, wiring harnesses or 3. GETTING STARTED3.1. Unpacking the M102E 3.2. Unpacking the M501-TRS3.2.1. M501-TRS Ventilation Clearance that the rack/enclosure itself is adequately ventilatedthe maximum operating temperature specification for the M102E 40C 3.3. Internal Layouts Figure 3-1 M102E Internal LayoutFigure 3-2 M501-TRS Internal Layout M501 3.4. Internal Pneumatic Flow of the M102E & the M501-TRSTable 3-1 TRS - SO2 Switching Valve Operating Modes M102E INSTRUMENT CHASSIS5 Amp 3.5. Rear Panel Layout for the M102E & M501-TRSAlarm Output Figure 3-4 M102E Rear Panel LayoutFigure 3-6 Analog Output Connector 3.6. Initial Setup3.6.1. Electrical Connections 3.6.1.1. M102E Analog Output ConnectionsM501-TRS PNEMATIC CONNECTERS 3.6.1.2. M501-TRS Alarm Output Connections3.6.2. Pneumatic Connections M102E PNEMATIC CONNECTERS102E MODELSPAN GAS The exhaust from the instrument needs to be vented outside theZERO AIR 3.6.2.1. Connections with Internal Valve Options InstalledGas Dilution Calibrator VENTVENT if input is pressurized Valve Option OptionPossible Warning Messages at Start-Up 3.7. Initial Operation3.7.1. Startup / Warm Up of the M102E 3.7.2. Functional Check of the M102EDisplay area High Alarm LED Low Alarm LED3.7.3. Startup / Warm Up of the M501-TRS Process Variable3.8. Initial Calibration Page OPTIONAL HARDWARE AND SOFTWARE 4. OPTIONAL HARDWARE AND SOFTWARE4.1. Rack Mount Kits Options 20a, 20b, 21, 22 4.2. Calibration Valves OptionsZero/Span Valve Operating States Table 4-2 IZS Valve Operating States M102E INSTRUMENT CHASSIS4.3. Additional Manuals 4.3.1. Printed Manuals P/N4.3.2. Addendum on CD Part number M102E OPERATING INSTRUCTIONS 5.1.1. M102E Analog Output Signals5. M102E OPERATING INSTRUCTIONS Figure 5-1 Analog Output Connector KeyTable 5-1 M102E gas Measurement Modes 5.2. SETUP - DIAG Using the Diagnostics Functions5.1.2. Setting the M102E Gas Measurement Mode 5.2.1. M102E Analog I/O Configuration5.3.2. M102E Ethernet Host Name 5.3. SETUP - COMM Setting Up the M102E’s Communication Ports5.2.2. M102E Test Channel Output 5.3.1. M102E ID CodeZERO 5.4. Remote Operation of the Analyzer5.4.1. Control Inputs CONTROL INSPAN 5.4.2. Using the M102E with a Hessen Protocol Network5 VDC Power Supply 5.4.2.1. M102E Hessen Protocol Gas ID ListTable 5-6 Default Hessen Status Bit Assignments 5.4.2.2. Setting Hessen Protocol Status FlagsTeledyne Instruments customer service 6. M501-TRS OPERATING INSTRUCTIONS6.1. Basic M501-TRS Controls DO NOT OPERATE WITHOUT THE COVER OF THE M501TS CONVERTER INSTALLEDTable 6-1 M501-TRS Temperature Controls and Definitions 6.2. To Display The Current TemperatureDO NOT SET THE TEMPERATURE HIGHER THAN 1050OC 6.3. To Manually Adjust the Converter Oven Temperature6.4. Autotune the Temperature Controller 6.4.1. Initiating the Autotune Processhas reached a stable, constant temperature 6.5. M501TRS Alarm Relay Adjustment 6.4.2. Aborting the Autotune Processnecessary to repeat the autotune procedure Page 7.2. M501-TRS Calibration 7. CALIBRATION PROCEDURESUSER NOTES 7.1. M102E CalibrationPage INSTRUMENT MAINTENANCE 8. INSTRUMENT MAINTENANCE8.1. Additional and Updated Maintenance Procedures 8.1.1.1. Predicting When the SO2 Scrubber Should Be Replaced8.1.1. Maintaining the SO2 Scrubber 8.1.1.3. Changing the SO2 Scrubber Material 8.1.1.2. Checking the Function of the SO2 ScrubberPage 9.1.2. SO2 Ultraviolet Fluorescence 9. THEORY OF OPERATION9.1. Measurement Principle 9.1.1. TRS ConversionFigure 9-1 UV Absorption in the M102E Reaction Cell SO2 + hv214nm ⎯⎯Ia⎯→ SO2SO2 * ⎯⎯→ SO2 + hv330nm Ia = I0 1 − exp− axSO2F = kSO2 kSO2 * ⎯⎯F⎯→ SO2 + hv330nmEquation9-6 9.3.1. Sample gas Flow 9.2. The UV Light Path9.3. Pneumatic Operation 9.2.1. UV Lamp Shutter & PMT OffsetSAMPLE CHAMBERUV Lamp PMT HOUSING9.4. Electronic Operation 9.4.1. Sensor ModuleSample Air 9.4.1.1. Sample Chamber9.4.1.2. Sample Chamber Heating Circuit Sample Air Housing Outlet O-Ring O-Ring Seal SealP-I-D CONTROLLER 9.4.2.1. Thermal Switch9.4.2. M501-TRS electronics HEATER9.4.2.2. Temperature Alarms and Alarm Output 10.1.1.1. M102E Warning Messages 10. TROUBLESHOOTING & REPAIR10.1.1. Fault Diagnosis with Warning Messages 10.1.1.2. M501-TRS Error Codes10.1.2. Fault Diagnosis with Test Functions Table 10-2 Test Functions - Possible Causes for Out-Of-Range Valuesfunctioning 10.3.1. Excessive noise 10.2. M501-TRS Trouble shooting10.3. Other Performance Problems 10.2.1. TRS Converter Not Heating10.4. Subsystem Checkout 10.4.1. Checking the Efficiency of the M501-TRS SO2 Scrubber10.4.2. Checking the Efficiency of the M501-TRS TRS Æ SO2 Converter 10.5.1. UV Lamp Adjustment and/or Replacement 10.5. Additional Repair ProceduresAlways grasp the main body of the lamp 10.5.1.1. Adjusting the UV Lamp Peaking the LampALWAYS wear UV-Protective, Safety Glasses when working with the UV Lamp AssemblyFigure 10-1 Shutter Assembly - Exploded View 10.5.1.2. Replacing the UV LampFigure 10-2 Disassembling the Shutter Assembly 10.5.2. Replacing the UV filter/lensFigure 10-3 PMT Assembly - Exploded View 10.5.3. Replacing the PMT, HVPS or TEC3. Remove the reaction cell assembly 10.5.4. M102E PMT Hardware Calibration FACTORY CAL Figure 10-4 Pre-Amplifier Board Layout 6. Locate the Preamp board see FigureNORM PMT value of twice the ppb value of the span gas EXAMPLE 10.5.5. Replacing the TRS Converter Heating TubeIf a reporting range other than 500 ppb is used in this procedure Adjust the PMT until NORM PMT equals 1280 mV ± 10 mVelement off 10.6. Manually Programming the M501-TRS Temperature ControllerThe ceramic bobbins at each end of the heater assembly are fragile Be Careful115V/60Hz with a set value of 1000C 10.6.1. Temperature Controller Primary Menu ParametersTable 10-3 - Temperature Controller - Primary Parameter Settings Table 10-4 - Temperature Controller - Primary Parameter Settings 10.7. Technical Assistance05514 Rev A1 M102E/M501 TRS APPENDIX A - Version Specific Software Documentation APPENDIX A - Version Specific Software DocumentationAPPENDIX A-2 Model 102E Setup Variables Available Via Serial I/O APPENDIX A-4 Model 102E Signal I/O DefinitionsAPPENDIX A-1 M102E Software Menu Trees, Revision A.2 SAMPLE SETUPENTER SETUP PASS APPENDIX A-1 M102E Software Menu Trees, Revision A.2CLR 1,3 SAMPLETEST MSG 1,2ACAL1 SO2, 111, REPORTEDTR2, 112, REPORTED SAMPLE ENTER SETUP PASS Figure A-4 Secondary Setup Menu DIAGDIAG MEASUREMODE APPENDIX A-2 Setup Variables For Serial I/O, Revision A.2Deleted Setup Variables for M102E Software Revision A.2 Table A-2Table A-6 Test Functions Deleted from M102 Software Revision A.2 setting of STABILFREQ and STABILSAMPLESAPPENDIX A-3 Warnings and Test Functions, Revision A.2 Table A-4 Warning Messages deleted from M102 Software Revision A.2APPENDIX A-4 M102E Signal I/O Definitions, Revision A.2 APPENDIX A-4 M102E Signal I/O Definitions, Revision A.2M102E/M501 TRSRear board primary MUX analog inputs Function APPENDIX A-5 M102E iDAS Functions, Revision A.2Trigger Event CTEMPWAPPENDIX A-5 M102E iDAS Functions, Revision A.2 A-10M101E Manual - P/N 04740120 Rev A APPENDIX B - M102E Spare Parts ListAPPENDIX B - M102E Spare Parts List Table B-1 M102E Spare Parts ListAddendum to M101E Manual - P/N 04740 Rev A INSTRUMENTS Warranty/Repair Questionnaire Model 102EM102E/M501 TRS TELEDYNEINSTRUMENTS Addendum to M101E Manual - P/N 04740 Rev A APPENDIX D - ELECTRONIC SCHEMATICSAPPENDIX D - ELECTRONIC SCHEMATICS Table D-1 List of Included Electronic Schematics05518 Rev A APPENDIX D - ELECTRONIC SCHEMATICSUser Notes M102E/M501 TRS