Fluke 1750 specifications Input Terminals and Controls

Page 14

1750

Operators Manual

 

 

Table 3. Input Terminals and Controls

 

 

 

Number

 

Description

 

 

A

Standard 120/240 V 50/60 Hz power input for Recorder power

 

 

 

B

Voltage indicator

 

 

Steady

Voltage is present, not overloaded

 

Blinking

Voltage overload (overload threshold 1100 V)

C

Five voltage measurement inputs

 

 

D

Removable Secure Digital (SD) memory card for transfer of large quantities of data. You can

 

remove the card from the Recorder, insert in the PDA and transfer data to the PC via the

 

“synchronize data” feature of the PDA. A USB SD card reader may also be used.

 

 

E

SD status indicator

 

Steady green

SD card inserted

 

Blinking

Busy - do not remove the card

F

Wireless Controller Status Indicator

 

Steady Blue

Wireless controller enabled but not communicating

 

Blinking

Wireless controller communicating

 

Off

Inactive

G

Ethernet port. TCP/IP via Ethernet is used to connect the Recorder to a PC for downloading

 

recorded data, and for configuring the Recorder when not using the PDA. This is the

 

recommended method of data transfer due to its speed and reliability.

 

 

 

H

Busy Indicator

 

 

Blinking

Network is busy

 

 

 

I

Link indicator

 

 

Off

No link

 

Steady

Link Present

 

Blinking

Communicating with PC

J

Current Measurement Terminals (5)

 

 

 

K

Current indicator

 

 

Steady

Current probe in normal range

 

Blinking

> 110% of probe range, use a higher range probe

 

Off

Current insufficient to measure

L

Power LED

 

 

Steady green

AC input power normal, instrument is not booted or locked

 

Green/Orange blinking Recording

 

Blinking red

Recorder running on UPS

M

Ground Terminal

 

 

Connecting the supplemental ground terminal and the line cord safety ground to different

 

ground potentials creates a ground loop that can damage the Recorder. To avoid this, use the

 

supplemental ground terminal only when no protective earth ground connection exists through

 

the line power cord. If there is any chance that a safety ground connection does exist through

 

the line power cord connection, make sure to connect the supplemental ground terminal only to

 

the ground system used by the ac receptacle that powers the Recorder. See Figure 3.

 

 

 

6

Image 14
Contents 1750 Limited Warranty and Limitation of Liability Table of Contents 1750 List of Tables 1750 List of Figures 1750 Power Recorder IntroductionContacting Fluke Safety Information XW WarningAccessories Symbols MeaningFeatures Fluke 1750 Power RecorderInput Terminals and Controls Applying the Front Panel Decal Applying the Front Panel DecalConnecting the Supplemental Ground Terminal Installing Power View on the PDA Charging the PDA BatteryInstalling the Software Language Selection PC System Requirement for Power AnalyzeInstalling Power Analyze Installing the Power Recorder at a Facility Work FlowInstalling the Recorder Connecting the Recorder to the WiringVerifying Your Connection Finishing UPPower Type Diagrams Managing Recorded DataExample Branch circuit at an outlet One-Phase IT No Neutral One-Phase Split Phase GND Three-Phase Delta Three-Phase IT Three-Phase High Leg Example a variant of power transformer winding type Element Delta ØC/L3 ØB/L2 Communicating with the Recorder Using Power View Connecting to the RecorderNavigating in Power View on the PDA Power View MenusIcons on the Menu Bar Menu Tree FlukeHome Screen Azd113.bmp Downloading Data and Erasing Memory Azd117.bmp Azd108s.bmp Azd109s.bmp Viewing Live Data Scope ScreenMeter Screen Phasor ScreenHarmonics Screen Trend ScreenSetting Up the Recorder Viewing PhasesSetup screen shows setup options Setting the Clock Probe DetectSetting the IP Address Adding a Measurement DescriptionConfigure Nominal Power Values Using Phase SwapSetting the Volts and Current Ratio Assign Recorder Name and PasswordAzd132.bmp Change Display Language and Phase Identifier Working with AnnotationsAzd137.bmp Using Image or Voice Annotations Cleaning and Maintenance Turn off the RecorderRegulatory Information for Wireless Communication Synchronization and Sampling General SpecificationsInput Specifications Voltage and Current MeasurementsTransient Voltage Impulse Power and PF MeasurementVoltage and Current Measurement Accuracy Dip Sag and Swell MeasurementsExternal Interface Specifications Environmental and Safety Specifications