HCD-C5
7. INITIAL SETTING OF ADJUSTMENT VALUE
Note:
Mode which sets the adjustment results recorded in the
If initial setting is performed, perform all adjustments again excluding the temperature compensation offset adjustment.
For details of the initial setting, refer to “4. Precautions for Adjustments” (See page 36) and execute the initial setting before the adjustment as re- quired.
Procedure:
1.Press the . “R” or > “R” button to display “ADJ CLEAR” (C28).
2.Press the ENTER/YES “R” button. “Complete!” will be displayed momentarily and initial setting will be executed, after which “ADJ CLEAR” (C28) will be displayed.
Procedure:
1.Press the . “R” or > “R” button to display “TEMP ADJUST” (C03).
2.Press the ENTER/YES “R” button to select the “TEMP ADJUST” mode.
3.“TEMP = [OK]” and the current temperature data will be displayed.
4.To save the data, press the ENTER/YES “R” button. When not saving the data, press the MENU/NO “R” button.
5.When the ENTER/YES “R” button is pressed, “TEMP = SAVE” will be displayed and turned back to “TEMP ADJUST” (C03) display then. When the MENU/NO “R” button is pressed, “TEMP ADJUST” (C03) will be displayed immediately.
Specified Value:
The “TEMP = ” should be within “E0 - EF”, “F0 - FF”, “00 - 0F”, “10 - 1F” and “20 - 2F”.
8.RECORDING AND DISPLAYING THE IOP INFORMATION
The IOP data can be recorded in the
Recording Procedure:
1.Press the . “R” or >“R” button to display “Iop Write” (C05), and press the ENTER/YES “R” button.
2.The display becomes “Ref=@@@.@” (@ is an arbitrary number) and the numbers which can be changed will blink.
3.Input the IOP value on the optical
To select the number : Press the .“R” or >“R” button.
To select the digit : Press two buttons of VOL - and CD Z simultaneously.
4.When the ENTER/YES “R” button is pressed, the display becomes “Measu=@@@.@” (@ is an arbitrary number).
5.As the adjustment results are recorded for the step 4 value. Leave it as it is and press the ENTER/YES “R” button.
6.“Complete!” will be displayed momentarily. The value will be recorded in the
Display Procedure:
1.Press the . “R” or > “R” button to display “Iop Read”(C26) and press the ENTER/YES “R” button.
2.“@@.@/##.#” is displayed and the recorded contents are displayed.
@@.@ indicates the IOP value on the optical
3.To end, press the MENU/NO “R” button to display “Iop Read” (C26).
9.TEMPERATURE COMPENSATION OFFSET ADJUSTMENT
Save the temperature data at that time in the
Note:
1.Usually, do not perform this adjustment.
2.Perform this adjustment in an ambient temperature of 22 ˚C to 28 ˚C. Per- form it immediately after the power is turned on when the internal tempera- ture of the unit is the same as the ambient temperature of 22 ˚C to 28 ˚C.
3.When D101 has been replaced, perform this adjustment after the tem- perature of this part has become the ambient temperature.
10.LASER POWER ADJUSTMENT
Before starting adjustment;
The laser power adjustment value changes depending upon the types of the optical
Check the type of the optical
Check the IOP value of the optical
Connection: | laser | ||
| |||
| power meter | ||
|
|
|
|
|
|
|
|
Optical
objective lens
digital voltmeter
BD (MD) board |
|
|
|
|
|
|
|
|
|
| |
CN105 pin 1 (I+3V) |
|
|
|
|
|
|
|
| + |
| |
|
|
| |||
CN105 pin 2 (IOP) |
|
|
| – | |
|
|
|
Procedure:
1.Insert the laser power meter probe into the disc insertion slot and set it on top of the objective lens of the optical
M “R” button to move the optical
Connect the digital voltmeter to CN105 pin 1 (I+3V) and CN105pin 2 (IOP) on the BD (MD) board.
2.Press the . “R” or > “R” button to display “LDPWR ADJUST” (C04).
(Laser power : For adjustment)
3.Press the ENTER/YES “R” button once to display “LD 0.9 mW $ ”.
4.Press the . “R” or > “R” button until the laser power meter reading matches with the specified value as described in the following table.
SPECIFIED VALUE | 0.85 to 0.91 mW | ||
|
| ||
0.90 to 0.96 mW | |||
|
Press the ENTER/YES “R” button after setting the range knob of the laser power meter, and save the adjustment results. (“LD SAVE $ ” will be displayed for a moment)
5. Then “LD 7.0 mW $ ” will be displayed.
40