7.6 OPTION (Option)
Up/down movement |
|
| Set by DATA |
|
|
| |||||
by MENU UP/DOWN |
|
| UP/DOWN |
|
|
| |||||
|
|
|
|
|
|
|
|
|
| ・Display of SCENE file |
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
| |
| <FILE A> |
| |||||||||
| ・Setting RS232C communication | 19200bps, 9600bps | |||||||||
|
|
|
|
|
|
|
|
|
| ||
|
| BAUD RATE | 19200bps | baud rate |
| ||||||
|
| ・Setting SCNE file | A, B, C | ||||||||
|
| SCENE FILE | A | ||||||||
|
|
|
| ||||||||
|
| TEMP.ALARM | ON | ・Setting of temperature warning | ON, OFF | ||||||
|
| CMG OFFSET | 0 |
|
|
| |||||
|
|
|
|
|
|
| |||||
|
|
|
|
|
|
|
|
|
| ・Electronic multiplication offset 0~100 | |
|
|
|
|
|
|
|
|
|
|
|
|
(1)Changing communication baud rate
q Using the [MENU UP] or [MENU DOWN] button, place “→” on BAUD RATE. w Using the [DATA UP] or [DATA DOWN] button, select 9600 bps or 19200 bps.
(2)Changing the setting of SCENE file
q Using the [MENU UP] or [MENU DOWN] button, place “→” on XCENE FILE. w Using the [DATA UP] or [DATA DOWN] button, select A, B or C.
(3)Changing the setting of temperature warning
This unit provides a function of detecting abnormal increase of internal temperature. When abnormal temperature increase is detected, the warning indicator “TEMP” appears on the display. Power consumption is lowered by stopping CCD cooling circuit.
The internal detection temperature is set to a temperature, as a guideline, at which the cabinet temperature is assumed to be approximately 70 to 75°C (158 to 167°F).
To avoid using this function in the practical environment, the function can be set to off. q Using the [MENU UP] or [MENU DOWN] button, place “→” on TEMP. ALARM. w Using the [DATA UP] or [DATA DOWN] button, select ON or OFF.
- 30 -