4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual
3.5Removal
1.REDUCE PROCESS PRESSURE TO 50 PSIG OR LESS. Shut off purge line, if used. Disconnect wiring connections if X1 is selected for Table IV in the MSG.
2.DO NOT STAND BEHIND THE TUBE WHEN PERFORMING THIS STEP. While holding the compression handle from turning, turn the support grip two turns counterclockwise. Loosen the compression handle until the process pressure pushes the cell and support tube assembly out to its internal stop. If necessary, pull out by hand until stop is reached.
3.Close the ball valve. If the valve does not close easily, make sure the support tube is pulled all the way back.
4.Completely loosen the compression handle to withdraw the cell.
3.6Electrical Connections
The terminal board connections for recorder or analyzer are given in the appropriate directions furnished with the measuring instrument.
When the cell assembly includes a
To avoid the possibility of ac
Cells are available with leadwire up to 20 feet as specified in the MSG assembly. For distances greater than 20 feet, use the required length of cable and a junction box, both listed in Section 4.3. For assemblies supplied with a Universal Head, Fig.
14 | 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual | 7/99 |