3M eg3, eg4 user manual Dosimeter screen

Page 30

24

Customizing set-up parameters in DMS

Dosimeter setting

The Edge general settings are used to setup the instrument with basic parameters. The eg4 model features two virtual dosimeters in one with two customized, user-defined settings. Please refer to the Edge dosimeter settings table below for an explanation of setup details.

1.Click on the Dosimeter tab from the Edge’s Instrument Configuration panel (see 1). To change the dosimeter settings, follow one (or all) of the selection options below.

Click on the checkbox to enable the dosimeter (LED) heartbeat, Dose indicator LED, and/or the Meter 1/Meter 2 Threshold settings.

Click in the circular button to select the response time (fast or slow), peak weighting (C or Z), and/or RMS Weighting (A or C).

To set the numeric fields either type in a value or use the arrows to enter appropriate values for the Meter1/Meter 2 settings (threshold, upper limit, exchange rate, criterion level, and RMS weighting).

2.When Meter 1 and Meter 2 settings are “enabled”, this will log (for eg4 model only) two different setups simultaneously. Typically, users will set the meter 1 and meter 2 with different regulatory or local standards (such as EU directive and OSHA).

3.To set the dosimeter settings, save and send to the Edge (for more details see “Saving and

sending configurations”).

Dosimeter settings explained in the Edge dosimeter settings (below).

1

Figure 3-6: Dosimeter screen

053-644, Rev E

Page 24

eg4 & eg3

Image 30
Contents Edge Turning off Display and keypad Introduction DosimetryGetting started Checking the equipment Turning on 3MTMDetection Mangement Software DMSDocking Noise studiesViews on the Edge Running studiesTechnical Support Appendix a Specifications StandardsStopping a study User InterfacePhysical attributes of EdgeDock1 and EdgeDock5 Contacting 3M Quest Technologies International customersINDEX………………………………………………………………………………… List of Tables Introduction Illustration of how the dosimeter operates and processes Range of sound Time history manufacturing Peak Measurement 121.20 dB View your results via QSP-II Setting-up your dosimeter parameters for your noise studiesRunning your studies Develop corrective actionsParameter setup screen for eg4 Getting started Powered on with measurement view displayed LED indicators explained Edge LED Explanation IndicatorEdge Keypad EdgeDock1 EdgeDock1 and EdgeDock5 components  Powering EdgeDock1 or EdgeDock5 Attaching charger adapter to EdgeDock1 and EdgeDock5 How to charge Docking the Edge with a single and 5-bay docksBackside of EdgeDock1 Setting-up the Edge via DMS Sample summary data from DMS reporting Logging Views on Explanation Dosimeter views and definitionsDose WeightingViews on Explanation Edge Response LevelDescriptors Variants Lmax variants Dosimeter measurement descriptors/variantsCommunicating connecting a docked Edge to a pc New Hardware wizard for Edge Dock Dosimeter screen Dosimeter fields explained for the Edge Edge display screen Edge security setting Edge auto run screen 10 Edge firmware update screen 11 Saving and sending Edge setups Quick Help List typical dosimetry session Quick help list for running a study How to attach the dosimeter Positioning/Attaching the Edge Checking battery power Battery indicators Calibrating the dosimeter Calibrating Reset On/off enter Contrast On/off/enter buttonContrast icon Checkmark icon Is the selection to  How to run a noise study Display screen while running How to stop a noise study Indicator Icon ExplanationIndicators i.e., run, stop, battery, and OL  Viewing measurements Measurement view on the Edge How to lock or secure  How to unlock Downloading Communicating between your pc and the EdgeDMS start screen Recent downloaded data  Selecting and viewing dataExample of Edge Report in DMS Panel layout view viewing the data in charts and graphs Technical Support Microphone and windscreen assembly  Attaching/Detaching the windscreen and microphoneHard Hat Attachment Kit Housing WeightSize Memory capacity LifeType OperatingLED Green indicator LED amber indicatorLED Red indicator Size of EdgeDock1Dosimeters Status indicatorsDisplayed Data RMS RangeSecurity Auto-RunLogging eg4 only Analog Board Edge Block Diagram Service Department and Technical Support 1 800 International customersBackground Noise, effects Combining Noise sources Crest FactorCriterion Level CL Dose Criterion TimeDecibel dB Exchange Rate ER also known as the Doubling RateLavg Average Level Free field setting microphoneHearing Conservation HC Hertz Hz Lep,dMaximum Level Lmax Leq Equivalent LevelLogging Measurement RangeRandom incidence microphone PEL Permissible Exposure LevelProjected Time Response TimeTWA Time Weighted Average ThresholdWeighting A, C, Z, etc Upper Limit ULWindscreen Dose = Index Edge DockQSP-II Running studies