3M eg4, eg3 user manual  Reset, On/off enter

Page 41

35

Overview of running a study

Resetting the dosimeter

The Edge can be reset through the dosimeter keypad or via QSP-II. The steps below outline how to reset your dosimeter.

Reset

1.

Ensure the meter is Off.

 

 

2.

Turn the meter on by pressing the on/off enter

button.

 

3.

When the splash screen appears, press and hold the run/stop

and on/off enter

 

buttons until the splash disappears.

 

 

Edge Splash screen

Figure 4-5: Splash screen

4.The Setup menu appears on the Edge. Press the up or down arrow to select the reset icon (symbol).

NOTE: Setup menu is used to reset the dosimeter, change the contrast, and return to the main view.

Reset icon

Checkmark icon Is the selection to exit the setup menu

Arrows and enter key Indicates the valid keys are selected

Figure 4-6: Resetting the Edge from the keypad

5. Press and hold the on/off enter button.

A 3, 2, 1 countdown will appear, and then a shaded bar will appear indicating the dosimeter is being reset.

6.Press the down arrow until the checkmark icon is selected (denoted by the arrows). Then,

press the on/off enter button to return to the measurement view screen.

053-644, Rev E

Page 35

eg4 & eg3

Image 41
Contents Edge Getting started Checking the equipment Turning on Introduction DosimetryTurning off Display and keypad 3MTMDetection Mangement Software DMSViews on the Edge Noise studiesDocking Running studiesStopping a study Appendix a Specifications StandardsTechnical Support User InterfaceINDEX………………………………………………………………………………… Physical attributes of EdgeDock1 and EdgeDock5Contacting 3M Quest Technologies International customers List of Tables Introduction Illustration of how the dosimeter operates and processes Range of sound Time history manufacturing Peak Measurement 121.20 dB Running your studies Setting-up your dosimeter parameters for your noise studiesView your results via QSP-II Develop corrective actionsParameter setup screen for eg4 Getting started Powered on with measurement view displayed Edge LED Explanation Indicator LED indicators explainedEdge Keypad EdgeDock1 EdgeDock1 and EdgeDock5 components Attaching charger adapter to EdgeDock1 and EdgeDock5  Powering EdgeDock1 or EdgeDock5Docking the Edge with a single and 5-bay docks  How to chargeBackside of EdgeDock1 Setting-up the Edge via DMS Sample summary data from DMS reporting Logging Dosimeter views and definitions Views on ExplanationViews on Explanation Edge WeightingDose Response LevelDosimeter measurement descriptors/variants Descriptors Variants Lmax variantsCommunicating connecting a docked Edge to a pc New Hardware wizard for Edge Dock Dosimeter screen Dosimeter fields explained for the Edge Edge display screen Edge security setting Edge auto run screen 10 Edge firmware update screen 11 Saving and sending Edge setups Quick help list for running a study Quick Help List typical dosimetry sessionPositioning/Attaching the Edge  How to attach the dosimeterBattery indicators  Checking battery powerCalibrating  Calibrating the dosimeterOn/off enter  ResetContrast icon Checkmark icon Is the selection to  ContrastOn/off/enter button Display screen while running  How to run a noise studyIndicators i.e., run, stop, battery, and OL  How to stop a noise studyIndicator Icon Explanation Measurement view on the Edge  Viewing measurements How to unlock  How to lock or secureCommunicating between your pc and the Edge  DownloadingDMS start screen  Selecting and viewing data Recent downloaded dataExample of Edge Report in DMS Panel layout view viewing the data in charts and graphs Technical Support  Attaching/Detaching the windscreen and microphone Microphone and windscreen assemblyHard Hat Attachment Kit Size HousingWeight Type LifeMemory capacity OperatingLED Red indicator LED amber indicatorLED Green indicator Size of EdgeDock1Displayed Data Status indicatorsDosimeters RMS RangeLogging eg4 only SecurityAuto-Run Analog Board Edge Block Diagram International customers Service Department and Technical Support 1 800Background Noise, effects Criterion Level CL Combining Noise sourcesCrest Factor Decibel dB Criterion TimeDose Exchange Rate ER also known as the Doubling RateHearing Conservation HC Hertz Hz Free field setting microphoneLavg Average Level Lep,dLogging Leq Equivalent LevelMaximum Level Lmax Measurement RangeProjected Time PEL Permissible Exposure LevelRandom incidence microphone Response TimeThreshold TWA Time Weighted AverageWindscreen Weighting A, C, Z, etcUpper Limit UL Dose = Edge Dock IndexRunning studies QSP-II