3M eg3, eg4 user manual  Calibrating the dosimeter

Page 40

34

Overview of running a study

Calibrating

3M Quest Technologies recommends to calibrate your dosimeter before you run your study(ies) and after your study(ies) to ensure highly accurate measurement results. The steps below address how to calibrate for either a pre or post calibration.

Calibrating the dosimeter

1.Remove the windscreen from the dosimeter by unscrewing it. (See Figure 6-1 for an illustration, page 50.)

2.Ensure the Edge is turned on and in the stop mode. (See “Turning on” page 8, and “Turning Off” page 9).

3.Slide the Edge into the mouth of the calibrator.

NOTE: ensure the adapter ring is inserted into the mouth of the calibrator before you slide the Edge into the cavity.

4.Turn on the calibrator and ensure parameters are set to 114 dB and 1000 Hz. (See calibration manual for operation details.)

“CAL” flashes

Figure 4-4: Calibrating

5.CAL” will flash on the display when detected.

6.Press the on/off enter button to trigger the calibration display.

7.To match the values on the calibration display, press repeatedly the up or down arrow in order to select the exact output on the calibrator (i.e., 114 dB).

8.When you have selected the appropriate dB level, press the on/off enter button.

A shaded bar will appear loading your calibration information. Then a summary calibration screen will appear indicating the db level with date and time displayed.

9. To return to the measurement view screen, press the on/off enter

.

10. To end the calibration steps, turn off the calibrator. Detach the dosimeter from the calibrator.

053-644, Rev E

Page 34

eg4 & eg3

Image 40
Contents Edge Introduction Dosimetry Getting started Checking the equipment Turning onTurning off Display and keypad 3MTMDetection Mangement Software DMSNoise studies Views on the EdgeDocking Running studiesAppendix a Specifications Standards Stopping a studyTechnical Support User InterfaceContacting 3M Quest Technologies International customers Physical attributes of EdgeDock1 and EdgeDock5INDEX………………………………………………………………………………… List of Tables Introduction Illustration of how the dosimeter operates and processes Range of sound Time history manufacturing Peak Measurement 121.20 dB Setting-up your dosimeter parameters for your noise studies Running your studiesView your results via QSP-II Develop corrective actionsParameter setup screen for eg4 Getting started Powered on with measurement view displayed LED indicators explained Edge LED Explanation IndicatorEdge Keypad EdgeDock1 EdgeDock1 and EdgeDock5 components  Powering EdgeDock1 or EdgeDock5 Attaching charger adapter to EdgeDock1 and EdgeDock5 How to charge Docking the Edge with a single and 5-bay docksBackside of EdgeDock1 Setting-up the Edge via DMS Sample summary data from DMS reporting Logging Views on Explanation Dosimeter views and definitionsWeighting Views on Explanation EdgeDose Response LevelDescriptors Variants Lmax variants Dosimeter measurement descriptors/variantsCommunicating connecting a docked Edge to a pc New Hardware wizard for Edge Dock Dosimeter screen Dosimeter fields explained for the Edge Edge display screen Edge security setting Edge auto run screen 10 Edge firmware update screen 11 Saving and sending Edge setups Quick Help List typical dosimetry session Quick help list for running a study How to attach the dosimeter Positioning/Attaching the Edge Checking battery power Battery indicators Calibrating the dosimeter Calibrating Reset On/off enterOn/off/enter button  ContrastContrast icon Checkmark icon Is the selection to  How to run a noise study Display screen while runningIndicator Icon Explanation  How to stop a noise studyIndicators i.e., run, stop, battery, and OL  Viewing measurements Measurement view on the Edge How to lock or secure  How to unlock Downloading Communicating between your pc and the EdgeDMS start screen Recent downloaded data  Selecting and viewing dataExample of Edge Report in DMS Panel layout view viewing the data in charts and graphs Technical Support Microphone and windscreen assembly  Attaching/Detaching the windscreen and microphoneHard Hat Attachment Kit Weight HousingSize Life TypeMemory capacity OperatingLED amber indicator LED Red indicatorLED Green indicator Size of EdgeDock1Status indicators Displayed DataDosimeters RMS RangeAuto-Run SecurityLogging eg4 only Analog Board Edge Block Diagram Service Department and Technical Support 1 800 International customersBackground Noise, effects Crest Factor Combining Noise sourcesCriterion Level CL Criterion Time Decibel dBDose Exchange Rate ER also known as the Doubling RateFree field setting microphone Hearing Conservation HC Hertz HzLavg Average Level Lep,dLeq Equivalent Level LoggingMaximum Level Lmax Measurement RangePEL Permissible Exposure Level Projected TimeRandom incidence microphone Response TimeTWA Time Weighted Average ThresholdUpper Limit UL Weighting A, C, Z, etcWindscreen Dose = Index Edge DockQSP-II Running studies