Quantum HD 9021 manual Method of USE

Page 10

ENGLISH

The light emitted by a lamp varies with the square of the current, so there may very well be variations in values during the test which the human eye does not detect.

When measuring temperature choose the probe most suited for the case: immersion, penetration or surface probes. It is sufficient to immerse the probe in the liquid of which you want to measure the temperature or place in contact with the surface; in the case of penetration probes, the tip of the probe must penetrate the block in which the temperature is to be measu- red.

METHOD OF USE

Although the probes are sturdy, they must be used with due care so as to avoid spoiling, scoring or breaking the filters or diffusors. Do not use them at temperatures higher than 50°C; be careful when there is a concentration of luminous beams or arc lights.

-Do not use the probes in the presence of corrosive gases or liquids or immerse them in liquids unless they are made for these purposes.

-Always use the most suitable probe for the measurement to be taken.

-Be careful with the range of use of the probe.

-Always clean the probes carefully after use.

-The instrument is resistant to water but it is not watertight, so if it should fall into the water, take it out immediately and check that no water has infiltrated.

– 32 –

Image 10
Contents Instructions Manual 22 23 24 Quantum Photo Radiometer Key for activating the display of input a Input B socket, DIN 8-pole connectorMAN key for choosing the working scale manually Key for activating the display of input BDing in C or F Lx/fcdCount of the time elapsed Data ShesAutomatic Seconds, minutes or hours Is being takenThreshold this key is used to DisabledIf a high value is to be set Decrease the set valueOFL HOW to MeasureErr Probe ConnectionMethod of USE LOW Battery Warning Data bit length Start bit MaintenanceSerial Output Stop bit no parityAnalog Output AD RS 232CBlock Diagram Layout of Components Calibrating the Instrument HD Series HD 9021 Probe CalibrationTechnical Features CharacteristicsMeasurements and measuring ranges Start bit Stop bit No parity Display LCD 12 mm, function and photo-radiometric symbolsOver ±0.3% ± 1C ±1 digit Correction Order CodeSer for cosine correction Probes Typical response curve of the probe is as followsHD 9021 RAD/C HD 9021 RAD/PARTypical response curve of the probe is as follows HD 9021 UVA Producing erythematous action in the fieldBlack light in the field Producing bactericide germicide action in the fieldHD 9021 UVB HD 9021 UVCLux lx HD 9021/CdLight flux Nit nitEN61000-4-2, EN61010-1 level SafetyVoltage variations