Quantum manual Calibrating the Instrument HD, Series HD 9021 Probe Calibration

Page 16

ENGLISH

CALIBRATING THE INSTRUMENT HD 9021

1)Open the instrument.

2)Connect the prods of a precision meter to the following pins on the input A (or B) connector: the positive to pin 7 (V+) and the negative to pin 6 (GND).

3)With the instrument and the meter switched on, turn the trimmer TR1 until the display on the meter indicates 4Vdc.

4)Insert the photometric simulator in the connector A (or B).

5)Turn the trimmer TR2 until the display on the instrument shows the same value as the simulator.

SERIES HD 9021 PROBE CALIBRATION

1)Fit the probe in the instrument.

2)Fit the probe in the photometric bench.

3)Turn the trimmer T1 of the amplification and calibration circuit in the pro- be connector until the value shown on the display coincides with the value of the reference instrument.

2

5

 

R3

R6

 

R8

C2

R10

 

 

 

 

 

 

 

 

 

DIN8P

 

U1

R1

HOLE

 

 

 

U2

 

8

TR1

 

A

C

C3

R2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

R9

R7

R5

R4

C1

 

 

 

 

T1

 

 

 

 

 

GUARANTEE

 

 

 

 

 

 

 

This instrument is strictly inspected before being sold. However if there should be any defect due to manufacture and/or transport, apply to the dealer from whom you bought the instrument.

The guarantee period is 2 (two) years from the date of purchase. During this period all defects found by us will be repaired free of charge, exclu- ding those due to incorrect use and careless handling. The probes are not covered by the guarantee, as they can be irreparably dama- ged after only a few minutes of incorrect use.

– 38 –

Image 16
Contents Instructions Manual 22 23 24 Quantum Photo Radiometer Input B socket, DIN 8-pole connector MAN key for choosing the working scale manuallyKey for activating the display of input a Key for activating the display of input BDing in C or F Lx/fcdCount of the time elapsed Data ShesAutomatic Seconds, minutes or hours Is being takenDisabled If a high value is to be setThreshold this key is used to Decrease the set valueHOW to Measure ErrOFL Probe ConnectionMethod of USE LOW Battery Warning Maintenance Serial OutputData bit length Start bit Stop bit no parity Analog Output AD RS 232CBlock Diagram Layout of Components Calibrating the Instrument HD Series HD 9021 Probe CalibrationTechnical Features CharacteristicsMeasurements and measuring ranges Start bit Stop bit No parity Display LCD 12 mm, function and photo-radiometric symbolsOver ±0.3% ± 1C ±1 digit Correction Order CodeSer for cosine correction Probes Typical response curve of the probe is as followsHD 9021 RAD/C HD 9021 RAD/PARTypical response curve of the probe is as follows Producing erythematous action in the field Black light in the fieldHD 9021 UVA Producing bactericide germicide action in the fieldHD 9021 UVB HD 9021 UVCHD 9021/Cd Light fluxLux lx Nit nitEN61000-4-2, EN61010-1 level SafetyVoltage variations