PROBES
The primary purpose of the probes is to detect and measure light in its various aspects; for this reason there are different probes available to suit the different requirements. Probes are required for measuring illuminance (lux,
7.34mm2. The probes are housed in a container of black anodized alumi- nium anticorodal UNI 9006/4; the correction filters are made of special glass, while the diffusors are in quartz or plastic material specially desi- gned for the purpose. The probe is complete with 2 metres of flexible cable and a DIN 45236
The probes are calibrated individually and are therefore interchan- geable.
HD 9021 PHOT/C
Photometric probe for measuring light, ILLUMINANCE, photopic correction filter complying with CIE, diffuser for cosine correction. Range 0÷200,000 lux (0÷200,000 foot candle). This is the probe for measuring light with correction according to the human eye; it is suitable for measure- ments in offices, schools, laboratories, places of entertainment, emergency exits, car parks, galleries, shops, shop windows, factories, etc., in all ligh- ting measurements.
The typical response curve of the probe is as follows:
1.00
0.89
0.78
0.67
0.56
0.44
0.33
0.22
0.11
0.00
380.0 | 420.0 | 460.0 | 500.0 | 540.0 | 580.0 | 620.0 | 660.0 | 700.0 | 740.0 | 780.0 |