Teledyne IGE 3000 ION Gauge Device Status 4-DIGIT Octal Word, Command Syntax

Page 18

8. Device Status

8.1.ION GAUGE DEVICE STATUS: 4-DIGIT OCTAL WORD

When requested to transmit its status the IGE-3000 responds with a 4-digit octal word.

FIGURE 7 -1 – DIGIT OCTAL WORD

bit 11: Serial Receiver Overload

bit 10: Main Board EEPROM Error.

bit 9: Communication Syntax Error.

bit 7 Emission Setting.

bit 8 Emission Setting.

bit 6: Filament Selection. (0=filament 1, 1=filament 2)

FIGURE 8- BITS 7 & 8 EMISSION CODE

BITS 7 & 8

BIT 8

BIT 7

NO EMISSIONS

0

0

0.01 Ma

0

1

0.1 mA

1

0

1.0 mA

1

1

bit 5: Filament Status. (0=non-operating, 1=operating)

bit 4: Degas (0=degas off, 1=degas on)

bit 3: Pressure less than (0.1e-9 Torr) Set if true.

bit 2: Pressure Greater than (1.0e-3 Torr) set if true.

bit 1: High Set point Alarm Set if pressure is greater than the high set point

bit 0: Low Set point Alarm Set if pressure is less than the low set point

8.2.Command Syntax

In the following examples of syntax codes, the special characters are explained:

The characters in square brackets [ ] represents a command string, either upper or lower case command characters accepted. All characters must follow each other in the string with no spaces or other characters.

The characters within wavy brackets { } contain choices for the appropriate command.

The characters within the symbols < > are the common abbreviations for the one digit ASCII control codes which they represent, (e.g. <CR> represents carriage return).

IGE 300

Page 18 of 21

Image 18
Contents Teledyne Hastings Manual Print History Table of Contents IGE Quick Start Information To Turn on the TUBE’S FilamentsGeneral Information SpecificationsGeneral Information IGE-3000 Mini Bayard/Alpert B/A TubeMaterials exposed to Vacuum Receiving Inspection InstallationInstallation MountingMounting B/A tube to IGE-3000 Electronics Module Connector Electrical Pin outsB/A Tube Installation EnvironmentalOperation Apply PowerOperation Front Panel OverviewOperating Modes Fil filament Mode DGas ModeFactor Mode SP set point ModeRemote Filament Control Theory of Operation Bayard Alpert B/A TubeRS485 Address Set Up GuideAdvance Setup Guide Baud RateData Bits/Parity Bits/Stop Bits Turnaround DelayRestore Eeprom Save EepromWiring for IGE-3000 RS232/485 Wiring for RS232/485 RS 232/RSRS232/RS485 Command Syntax Device StatusION Gauge Device Status 4-DIGIT Octal Word Interrogation Commands Parameter Modification CommandsCommand Description Format Sample Response Warranty Repair Policy Non-Warranty Repair PolicyDrawings & Schematics
Related manuals