Teledyne IGE 3000 instruction manual Interrogation Commands, Parameter Modification Commands

Page 19

When entering more than one command in the same data string, they must be separated by a comma.

All command strings must be followed by the terminator character (carriage return <CR>, also known as ENTER).

When a lower case character is present in an example it represents an option.

Character Description Valid Inputs:

a RS-485 Address (hexadecimal 0-9, A-F) 01 - DF

m Most Significant Digit Of Mantissa 1 - 9 d Decimal Digit 0 - 9

e Exponent 0 - 5

u Unit Of Pressure T, M, or P

, Command Separator (comma) N/A

<CR> Command Terminator (carriage return) N/A

8.3.Interrogation Commands

Command Description

Format

Sample Response

Transmit K Factor

K<CR>

K Factor: 1.0000+01

Transmit Electrometer Output

W<CR>

V1 Ave: 1.2765e+00

Transmit Fil Status

F<CR>

Filament #1 on High Voltage on

Transmit Averaged Pressure

P<CR>

Pa: 1.23456e+0 Torr<CR>

Transmit RS-485 Address

A<CR>

Multidrop Address: 01<CR>

Transmit High Set point

H<CR>

Hi: 1.00000e+1 Torr<CR>

Transmit Low Set point

L<CR>

Lo: 1.00000e-2 Torr<CR>

Transmit Device Status

S<CR>

00044<CR>

Transmit Turnaround Delay

T<CR>

Comm Delay: 6<CR>

Transmit Selected Units

U<CR>

Torr<CR>

Transmit Software Version #

V<CR>

IGE-3000 Vxxxx

Transmit Filament F<CR>

f1<CR>

 

Transmit Emission Setting

I<CR>

Emission: 0.01 mA<CR> (note:

 

 

Filament must be on)

Transmit Degas Time

D<CR>

Remaining Degas Time: 54

 

 

minutes<CR>

8.4.Parameter Modification Commands

Command Description

Format

Sample Response

Modify High Setpoint

H={m.dd}E{+e}<CR> 1.00000e-9 to 9.99999e+9

Modify Low Setpoint

L={m.dd}E{-e}<CR>

1.00000e-9 to 9.99999e+9

Modify RS-485 Address

*{aa}A={aa}<CR>

1 To DF (Hexadecimal)

Modify Turnaround Delay

*{aa}T={dd}<CR>

0 To 255 (Decimal)

Modify Emission

E={m.d}E{-e}<CR>

* must be exact

Modify Degas Time

R={dd}<CR>

0 To 5 (Decimal)

Turn Gauge/Filament on F1 or F2

F1<CR>

NA

Turn Gauge/Filament off

F0<CR>

NA

Escape

<Esc>

Reset Command Buffer (ignore prior

 

 

Input)

IGE 300

Page 19 of 21

Image 19
Contents Teledyne Hastings Manual Print History Table of Contents IGE To Turn on the TUBE’S Filaments Quick Start InformationIGE-3000 Mini Bayard/Alpert B/A Tube SpecificationsGeneral Information General InformationMaterials exposed to Vacuum Mounting InstallationInstallation Receiving InspectionEnvironmental Connector Electrical Pin outsB/A Tube Installation Mounting B/A tube to IGE-3000 Electronics ModuleFront Panel Overview Apply PowerOperation OperationOperating Modes SP set point Mode DGas ModeFactor Mode Fil filament ModeRemote Filament Control Bayard Alpert B/A Tube Theory of OperationBaud Rate Set Up GuideAdvance Setup Guide RS485 AddressSave Eeprom Turnaround DelayRestore Eeprom Data Bits/Parity Bits/Stop BitsRS 232/RS Wiring for IGE-3000 RS232/485 Wiring for RS232/485RS232/RS485 Device Status Command SyntaxION Gauge Device Status 4-DIGIT Octal Word Parameter Modification Commands Interrogation CommandsCommand Description Format Sample Response Non-Warranty Repair Policy Warranty Repair PolicyDrawings & Schematics