•When entering more than one command in the same data string, they must be separated by a comma.
•All command strings must be followed by the terminator character (carriage return <CR>, also known as ENTER).
•When a lower case character is present in an example it represents an option.
•Character Description Valid Inputs:
a RS-485 Address (hexadecimal 0-9, A-F) 01 - DF
m Most Significant Digit Of Mantissa 1 - 9 d Decimal Digit 0 - 9
e Exponent 0 - 5
u Unit Of Pressure T, M, or P
, Command Separator (comma) N/A
<CR> Command Terminator (carriage return) N/A
8.3.Interrogation Commands
Command Description | Format | Sample Response |
Transmit K Factor | K<CR> | K Factor: 1.0000+01 |
Transmit Electrometer Output | W<CR> | V1 Ave: 1.2765e+00 |
Transmit Fil Status | F<CR> | Filament #1 on High Voltage on |
Transmit Averaged Pressure | P<CR> | Pa: 1.23456e+0 Torr<CR> |
Transmit | A<CR> | Multidrop Address: 01<CR> |
Transmit High Set point | H<CR> | Hi: 1.00000e+1 Torr<CR> |
Transmit Low Set point | L<CR> | Lo: |
Transmit Device Status | S<CR> | 00044<CR> |
Transmit Turnaround Delay | T<CR> | Comm Delay: 6<CR> |
Transmit Selected Units | U<CR> | Torr<CR> |
Transmit Software Version # | V<CR> | |
Transmit Filament F<CR> | f1<CR> |
|
Transmit Emission Setting | I<CR> | Emission: 0.01 mA<CR> (note: |
|
| Filament must be on) |
Transmit Degas Time | D<CR> | Remaining Degas Time: 54 |
|
| minutes<CR> |
8.4.Parameter Modification Commands
Command Description | Format | Sample Response |
Modify High Setpoint | H={m.dd}E{+e}<CR> | |
Modify Low Setpoint | ||
Modify | *{aa}A={aa}<CR> | 1 To DF (Hexadecimal) |
Modify Turnaround Delay | *{aa}T={dd}<CR> | 0 To 255 (Decimal) |
Modify Emission | * must be exact | |
Modify Degas Time | R={dd}<CR> | 0 To 5 (Decimal) |
Turn Gauge/Filament on F1 or F2 | F1<CR> | NA |
Turn Gauge/Filament off | F0<CR> | NA |
Escape | <Esc> | Reset Command Buffer (ignore prior |
|
| Input) |
IGE 300 | Page 19 of 21 |