Teledyne IGE 3000 instruction manual Remote Filament Control

Page 13

The IGE-3000’s filament may be controlled remotely via the D power supply connector by wiring relay contacts or a transistor switch across pins 8 and 9, (refer to section 2.5 /Figure 1 and Figure 3 below). Pin 8 is digital ground and pin 9 is the filament control line. The IGE filament will turn on when pin 9 is pulled low or grounded. This may be accomplished either by relay contacts or by utilizing a transistor switch. A contact closure connected across “D” connector pin 8 and pin 9 must be such that it closes when the IGE- 3000 is to be turned on. The transistor circuit connected across “D” connector pin 8 and pin 9 should be such that it conducts and provides a ground path when the IGE-3000 is to be turned on.

FIGURE 3- REMOTE FILAMENT CONTROL

 

CONTROLLING

 

 

 

 

 

 

 

 

 

 

VACUUM

 

IGE 3000

 

 

 

GAUGE

 

 

 

 

"D" POWER CONNECTOR

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

K1

 

 

 

 

PIN 9

 

REMOTE START

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

K

 

 

 

 

 

 

 

 

 

 

 

 

RELAY INTERFACE

 

 

 

 

 

 

 

 

 

PIN 8

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CONTROLLING

 

IGE 3000

 

 

VACUUM

 

 

 

"D" POWER CONNECTOR

 

 

 

GAUGE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

PIN 9

 

REMOTE START

 

 

 

 

Q1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TRANSISTOR

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

INTERFACE

 

 

 

 

 

 

 

 

 

 

PIN 8

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

NOTE: Manual operation is unaffected by using remote control. (i.e. Filament can be turned off manually if turned on via remote control and filament can be turned on manually, after being turned off via remote control).

IGE 300

Page 13 of 21

Image 13
Contents Teledyne Hastings Manual Print History Table of Contents IGE To Turn on the TUBE’S Filaments Quick Start InformationGeneral Information SpecificationsGeneral Information IGE-3000 Mini Bayard/Alpert B/A TubeMaterials exposed to Vacuum Installation InstallationReceiving Inspection MountingB/A Tube Installation Connector Electrical Pin outsMounting B/A tube to IGE-3000 Electronics Module EnvironmentalOperation Apply PowerOperation Front Panel OverviewOperating Modes Factor Mode DGas ModeFil filament Mode SP set point ModeRemote Filament Control Bayard Alpert B/A Tube Theory of OperationAdvance Setup Guide Set Up GuideRS485 Address Baud RateRestore Eeprom Turnaround DelayData Bits/Parity Bits/Stop Bits Save EepromRS 232/RS Wiring for IGE-3000 RS232/485 Wiring for RS232/485RS232/RS485 Device Status Command SyntaxION Gauge Device Status 4-DIGIT Octal Word Parameter Modification Commands Interrogation CommandsCommand Description Format Sample Response Non-Warranty Repair Policy Warranty Repair PolicyDrawings & Schematics
Related manuals
Manual 21 pages 27.57 Kb