DANGER HIGH VOLTAGE

High voltages capable of causing injury or death are present within the IGE 3000 Avoid touching the connector sockets, tube pins and all internal circuitry within the IGE 3000. SERVICE MUST ONLY BE PERFORMED BY QUALIFIED TELEDYNE INSTRUMENTS PERSONNEL.

DANGER

Do not turn on the IGE 3000 if there is any danger of explosion from explosive or combustible gasses or gas mixtures. Ionization gauge tube filaments operate at a temperature sufficiently high enough to be considered an ignition source.

DANGER

Safe operation of ion producing equipment, including the IGE 3000, requires grounding of its power supply and the vacuum chamber. LETHAL VOLTAGES may be established under some operating conditions unless grounding is provided.

DANGER

Ion producing equipment such as the IGE 3000, from any manufacturer, may, under some conditions provide sufficient conduction via plasma to couple a high voltage electrode to the vacuum chamber. If conductive parts of the chamber are not grounded, they may attain a potential near that of the high voltage electrodes during this coupling. Potentially fatal electrical shock could then occur because of the high voltage between these vacuum chamber parts and ground. Grounding, though simple, is imperative! Ensure that the ground circuits are correctly utilized, both on your IGE 3000 power supplies and your vacuum chambers, regardless of the manufacturer, for this phenomenon is not peculiar to the IGE 3000.

CAUTION

Ensure that BA tube filaments are not turned on unless vacuum IS LESS THAN 10-2Torr.

TURNING ON filaments at atmosphere MAY result in damage to the BA tube.

CAUTION

Before initiating Degas, vacuum must be 10-7Torr or less.

Performing Degas at Higher Pressures Could Damage the B/A Tube

IGE 300

Page 4 of 21

Image 4
Contents Teledyne Hastings Manual Print History Table of Contents IGE Quick Start Information To Turn on the TUBE’S FilamentsSpecifications General InformationGeneral Information IGE-3000 Mini Bayard/Alpert B/A TubeMaterials exposed to Vacuum Installation InstallationReceiving Inspection MountingConnector Electrical Pin outs B/A Tube InstallationMounting B/A tube to IGE-3000 Electronics Module EnvironmentalApply Power OperationOperation Front Panel OverviewOperating Modes DGas Mode Factor ModeFil filament Mode SP set point ModeRemote Filament Control Theory of Operation Bayard Alpert B/A TubeSet Up Guide Advance Setup GuideRS485 Address Baud RateTurnaround Delay Restore EepromData Bits/Parity Bits/Stop Bits Save EepromRS 232/RS Wiring for IGE-3000 RS232/485 Wiring for RS232/485RS232/RS485 Device Status Command SyntaxION Gauge Device Status 4-DIGIT Octal Word Parameter Modification Commands Interrogation CommandsCommand Description Format Sample Response Warranty Repair Policy Non-Warranty Repair PolicyDrawings & Schematics