7. RS 232/RS 485

7.1.RS232/RS485

The IGE-3000 is designed to communicate with external devices by either RS232 or RS485 via the two RJ-11 connectors. Refer to section 3.2 Figure 2 for connector locations. RS232 via the left connector on and the RS485 is via either connector. Software capable of communicating with the IGE-3000 is available upon request.

Refer to Figure 7 below for pin out and wiring information.

7.1.1.Wiring for IGE-3000 RS232/485

7.1.2.Wiring for RS232/485

 

RS 485 half duplex

 

 

RS 485 full duplex

Pin #1

T+/R+

Pin #1

 

T+

 

 

 

 

 

 

Pin #2

 

T-/R-

Pin #2

 

T-

Pin #3

GND

Pin #3

 

GND

 

 

 

 

 

 

Pin #4

GND

Pin #4

 

GND

 

 

 

 

 

 

Pin #5

n/a

Pin #5

 

R-

 

 

 

 

 

 

Pin #6

 

n/a

Pin #6

 

R+

IGE 300

Page 17 of 21

Image 17
Contents Teledyne Hastings Manual Print History Table of Contents IGE To Turn on the TUBE’S Filaments Quick Start InformationGeneral Information SpecificationsGeneral Information IGE-3000 Mini Bayard/Alpert B/A TubeMaterials exposed to Vacuum Installation InstallationReceiving Inspection MountingB/A Tube Installation Connector Electrical Pin outsMounting B/A tube to IGE-3000 Electronics Module EnvironmentalOperation Apply PowerOperation Front Panel OverviewOperating Modes Factor Mode DGas ModeFil filament Mode SP set point ModeRemote Filament Control Bayard Alpert B/A Tube Theory of OperationAdvance Setup Guide Set Up GuideRS485 Address Baud RateRestore Eeprom Turnaround DelayData Bits/Parity Bits/Stop Bits Save EepromRS232/RS485 Wiring for IGE-3000 RS232/485 Wiring for RS232/485RS 232/RS ION Gauge Device Status 4-DIGIT Octal Word Command SyntaxDevice Status Command Description Format Sample Response Interrogation CommandsParameter Modification Commands Non-Warranty Repair Policy Warranty Repair PolicyDrawings & Schematics
Related manuals
Manual 21 pages 27.57 Kb