Teledyne IGE 3000 Connector Electrical Pin outs, B/A Tube Installation, Environmental

Page 9

3.5.“D” Connector Electrical Pin outs

The figure below indicates all pin outs.

FIGURE 1- IGE-3000 “D” CONNECTOR

NOTE: Set Point High, Set Point Low and Remote

Start, use Digital ground Pin 8 as signal return path.

3.6.B/A Tube Installation

The Bayard/Alpert (B/A) tube may be installed in any orientation. Although the transducer tube is rugged and will perform well in many harsh environments, the tube should be installed in a clean and careful manner. The B/A tube is configured with the vacuum fitting requested.

CAUTION

The electronics unit is calibrated when married to the B/A tube. Avoid interchanging B/A tubes with different electronics modules since this may affect accuracy and will void calibration.

3.7.Mounting B/A tube to IGE-3000 Electronics Module

The B/A tube plugs into the IGE-3000 via the 12 pin circular connector located inside the lower base housing of the IGE-3000 electronics module. The B/A tube is keyed to ensure proper connections. Slightly rotate the electronics unit until the keyway of the connector aligns with the key on the B/A tube, and then snugly plug the electronics module to the B/A tube. Tighten the set screw located on the base housing to the B/A tube. The set screw secures the tube to the electronics module and also ensures proper grounding of the B/A tube to the IGE-3000 base housing.

3.8.Environmental:

Indoor use

Operating temperature range from 0° to 40° C

Maximum relative humidity: 80% for temperatures up to 31° C decreasing linearly to 50% Relative humidity at 40° C.

IGE 300

Page 9 of 21

Image 9
Contents Teledyne Hastings Manual Print History Table of Contents IGE To Turn on the TUBE’S Filaments Quick Start InformationGeneral Information SpecificationsGeneral Information IGE-3000 Mini Bayard/Alpert B/A TubeMaterials exposed to Vacuum Installation InstallationReceiving Inspection MountingB/A Tube Installation Connector Electrical Pin outsMounting B/A tube to IGE-3000 Electronics Module EnvironmentalOperation Apply PowerOperation Front Panel OverviewOperating Modes Factor Mode DGas ModeFil filament Mode SP set point ModeRemote Filament Control Bayard Alpert B/A Tube Theory of OperationAdvance Setup Guide Set Up GuideRS485 Address Baud RateRestore Eeprom Turnaround DelayData Bits/Parity Bits/Stop Bits Save EepromWiring for IGE-3000 RS232/485 Wiring for RS232/485 RS 232/RSRS232/RS485 Command Syntax Device StatusION Gauge Device Status 4-DIGIT Octal Word Interrogation Commands Parameter Modification CommandsCommand Description Format Sample Response Non-Warranty Repair Policy Warranty Repair PolicyDrawings & Schematics
Related manuals
Manual 21 pages 27.57 Kb