3.5.“D” Connector Electrical Pin outs
The figure below indicates all pin outs.
FIGURE 1- IGE-3000 “D” CONNECTOR
NOTE: Set Point High, Set Point Low and Remote
Start, use Digital ground Pin 8 as signal return path.
3.6.B/A Tube Installation
The Bayard/Alpert (B/A) tube may be installed in any orientation. Although the transducer tube is rugged and will perform well in many harsh environments, the tube should be installed in a clean and careful manner. The B/A tube is configured with the vacuum fitting requested.
CAUTION
The electronics unit is calibrated when married to the B/A tube. Avoid interchanging B/A tubes with different electronics modules since this may affect accuracy and will void calibration.
3.7.Mounting B/A tube to IGE-3000 Electronics Module
The B/A tube plugs into the
3.8.Environmental:
Indoor use
Operating temperature range from 0° to 40° C
Maximum relative humidity: 80% for temperatures up to 31° C decreasing linearly to 50% Relative humidity at 40° C.
IGE 300 | Page 9 of 21 |