Teledyne 3190Z operating instructions Contact ID for Failsafe Relay Operation

Page 23

Trace Oxygen Analyzer

Installation

 

 

 

Alarm Relays: The three alarm circuit connectors are screw terminals for making connections to internal alarm relay contacts. There is one set of contacts for each type of alarm. Contacts are Form C, with normally open and normally closed contact connections capable of switching up to 0.5A at 125 VAC into a resistive load.

The alarm relay circuits are designed for failsafe operation, meaning the relays are energized during normal operation. If power fails the relays de-energize (alarms activated).

The contact connections are indicated diagrammatically on the rear panel as normally closed (NC), common (C), and normally open (NO). Figure 3-3 explains how these act in failsafe operation.

The specific descriptions for each type of alarm are as follows:

HI Alarm Configured as high alarm (actuates when concentration is above threshold). Can be set anywhere within the full range of the analyzer (1- 1,000 ppm), but must be set ABOVE the threshold set for the LO Alarm.

LO Alarm Configured as low alarm (actuates when concentration is below threshold). within the full range of the analyzer (1-1,000 ppm), but must be set BELOW the threshold set for the HI Alarm.

Sensor Fail Not active in this product

Figure 3-3: Contact ID for FAILSAFE Relay Operation

Teledyne Analytical Instruments

15

Image 23
Contents Teledyne Analytical Instruments Trace Oxygen AnalyzerCopyright 2009 Teledyne Instruments/ Analytical Instruments Model 3190ZSafety Messages Model 3190Z Table of Contents Maintenance Combustible GAS Usage Viii Trace Oxygen Analyzer Introduction Main Features of the AnalyzerIntroduction OverviewSpan Front Panel DescriptionIntroduction Model 3190Z Measurement Parameters Readout Rear Panel DescriptionMeter Readout Rear Panel AC and DC versions ƒ Power ConnectionAlarm Page Trace Oxygen Analyzer Operational Theory Operational TheoryIntroduction Zirconium Oxide SensorsGeneral ElectronicsOperational Theory Model 3190Z Effect of PressureBlock Diagram of the Signal Processing Electronics Signal ProcessingOperational Theory Model 3190Z Unpacking the Analyzer InstallationTrace Oxygen Analyzer Installation Location and Mounting Installation Model 3190ZControl Unit Installation External Probe InstallationPower switch is located below the fuse receptacle Rear Panel Electrical Connectors for AC and DC Units Contact ID for Failsafe Relay Operation Gas Connections Installation ChecklistTrace Oxygen Analyzer Operation OperationOperation Model 3190Z Using the Function and Data Entry ButtonsSetting the Analysis Ranges HI Range Setting the Alarm SetpointsHI Alarm LO AlarmCalibration Sensor Fail AlarmProcedure Selecting a Fixed Range or AutorangingLED Display on StartUp Page Replacing the Fuse MaintenanceTrace Oxygen Analyzer Maintenance AC Powered UnitsWhen to Replace a Sensor Sensor Installation or ReplacementMaintenance Model 3190Z DC Powered UnitsRange ID 0-10 VDC SpecificationsAppendix Trace Oxygen Analyzer AppendixAppendix Model 3190Z Spare Parts ListMiscellaneous Drawing List