Teledyne 3190Z Setting the Alarm Setpoints, HI Range, LO Range, HI Alarm, LO Alarm

Page 27

Trace Oxygen Analyzer

Operation

 

 

 

4.3.1 HI Range

Setting the HI Range full-scale value defines the LEAST sensitive analysis range to be used. To set the HI Range:

1.Press the SET HI RANGE function button once.

2.Immediately (within 5 seconds) press either the Δ or button to raise or lower the displayed value as required, until the display reads the desired full-scale concentration.

4.3.2LO Range

Setting the LO Range full-scale value defines the MOST sensitive range to be used. To set the LO Range:

1.Press the SET LO RANGE function button once.

2.Immediately (within 5 seconds) press either the Δ or button to raise or lower the displayed value as required, until the display reads the desired full-scale concentration.

4.4Setting the Alarm Setpoints

The alarm set-points can be adjusted over the full range of the analyzer (0-1,000 ppm oxygen content). The set-point values are expressed in ppm only.

Note: The HI Alarm set-point MUST be set at a higher concentration than the LO Alarm set-point.

4.4.1 HI Alarm

Setting the HI Alarm sets the value ABOVE which the HI Alarm will activate. To Set the HI Alarm:

1.Press the SET HI ALARM function button once.

2.Within 5 seconds press either the Δ or button to raise or lower the displayed value as required, until the display reads the desired concentration.

4.4.2LO Alarm

Setting the LO Alarm sets the value BELOW which the LO alarm will activate. To set the LO Alarm:

Teledyne Analytical Instruments

19

Image 27
Contents Teledyne Analytical Instruments Trace Oxygen AnalyzerCopyright 2009 Teledyne Instruments/ Analytical Instruments Model 3190ZSafety Messages Model 3190Z Table of Contents Maintenance Combustible GAS Usage Viii Trace Oxygen Analyzer Introduction Main Features of the AnalyzerIntroduction OverviewFront Panel Description Introduction Model 3190ZSpan Rear Panel Description Meter ReadoutMeasurement Parameters Readout Rear Panel AC and DC versions ƒ Power ConnectionAlarm Page Trace Oxygen Analyzer Operational Theory Operational TheoryIntroduction Zirconium Oxide SensorsGeneral ElectronicsOperational Theory Model 3190Z Effect of PressureBlock Diagram of the Signal Processing Electronics Signal ProcessingOperational Theory Model 3190Z Installation Trace Oxygen Analyzer InstallationUnpacking the Analyzer Location and Mounting Installation Model 3190ZControl Unit Installation External Probe InstallationPower switch is located below the fuse receptacle Rear Panel Electrical Connectors for AC and DC Units Contact ID for Failsafe Relay Operation Gas Connections Installation ChecklistTrace Oxygen Analyzer Operation OperationUsing the Function and Data Entry Buttons Setting the Analysis RangesOperation Model 3190Z HI Range Setting the Alarm SetpointsHI Alarm LO AlarmCalibration Sensor Fail AlarmProcedure Selecting a Fixed Range or AutorangingLED Display on StartUp Page Replacing the Fuse MaintenanceTrace Oxygen Analyzer Maintenance AC Powered UnitsWhen to Replace a Sensor Sensor Installation or ReplacementMaintenance Model 3190Z DC Powered UnitsRange ID 0-10 VDC SpecificationsAppendix Trace Oxygen Analyzer AppendixAppendix Model 3190Z Spare Parts ListMiscellaneous Drawing List