Teledyne 3190Z Specifications, Trace Oxygen Analyzer Appendix, Range ID 0-10 VDC

Page 33

Trace Oxygen Analyzer

Appendix

 

 

 

Appendix

A.1 Specifications

Ranges: Two user selectable ranges can be set between 0-10 ppm and 0-1,000 ppm oxygen. Default ranges are 0-100 ppm and 0-1,000 ppm oxygen

Signal Output: Voltage: 0–10 VDC, negative ground Current: 4-20 mA, negative ground

Range ID: 0-10 VDC

Display: Light-emitting diode (LED) display

Alarms: One high alarm relay, adjustable; one low alarm relay, adjustable; one sensor failure relay. All alarms failsafe.

System Operating Temp: 0 - 50°C

Accuracy: ±2% of full scale at constant temperature ±5% of full scale over 0-50°C range once temperature equilibrium is reached (at 3% and 5% user defined ranges).

Response Time: 90% in less than 20 seconds at 25°C

System Power: AC (100-240 VAC, 47/440 Hz), or DC

Requirements: (10-36 VDC); user specified

System Enclosure: Panel Mount: 2.81" H × 6.0" W × 2.87" D (71.4 mm × 152.4 mm × 72.9 mm).

Face Plate: 3.75" H × 7.0" W (95.3 mm H × 177.8 mm W). Faceplate rated to NEMA-4

Sensor Type: Zirconium Oxide

Analysis Unit: 4.0" H × 6.0" W × 3.25" D (101.6 mm × 152.4 mm × 82.6 mm)

Teledyne Analytical Instruments

25

Image 33
Contents Teledyne Analytical Instruments Trace Oxygen AnalyzerCopyright 2009 Teledyne Instruments/ Analytical Instruments Model 3190ZSafety Messages Model 3190Z Table of Contents Maintenance Combustible GAS Usage Viii Introduction Main Features of the AnalyzerOverview Trace Oxygen Analyzer IntroductionFront Panel Description Introduction Model 3190ZSpan Rear Panel Description Meter ReadoutMeasurement Parameters Readout Rear Panel AC and DC versions ƒ Power ConnectionAlarm Page Introduction Operational TheoryZirconium Oxide Sensors Trace Oxygen Analyzer Operational TheoryOperational Theory Model 3190Z ElectronicsEffect of Pressure GeneralBlock Diagram of the Signal Processing Electronics Signal ProcessingOperational Theory Model 3190Z Installation Trace Oxygen Analyzer InstallationUnpacking the Analyzer Control Unit Installation Installation Model 3190ZExternal Probe Installation Location and MountingPower switch is located below the fuse receptacle Rear Panel Electrical Connectors for AC and DC Units Contact ID for Failsafe Relay Operation Gas Connections Installation ChecklistTrace Oxygen Analyzer Operation OperationUsing the Function and Data Entry Buttons Setting the Analysis RangesOperation Model 3190Z HI Alarm Setting the Alarm SetpointsLO Alarm HI RangeProcedure Sensor Fail AlarmSelecting a Fixed Range or Autoranging CalibrationLED Display on StartUp Page Trace Oxygen Analyzer Maintenance MaintenanceAC Powered Units Replacing the FuseMaintenance Model 3190Z Sensor Installation or ReplacementDC Powered Units When to Replace a SensorAppendix SpecificationsTrace Oxygen Analyzer Appendix Range ID 0-10 VDCAppendix Model 3190Z Spare Parts ListMiscellaneous Drawing List