Teledyne 3190Z operating instructions Block Diagram of the Signal Processing Electronics

Page 17

Trace Oxygen Analyzer

Operational Theory

 

 

 

2.3.2 Signal Processing

Figure 2-4 is a block diagram of the signal processing electronics described below.

Figure 2-4: Block Diagram of the Signal Processing Electronics

In the presence of oxygen, the sensor(s) produce a current. A current to voltage amplifier (I–E AMPL) converts this current to a voltage.

The output from the I-E amplifier and subsequent buffer circuit (not shown) is sent to an analog to digital converter (ADC), and the resulting digital concentration signal is sent to the micro controller.

The digital concentration signal along with input from the front panel buttons (KEYBOARD) is processed by the micro controller, and appropriate output signals are directed to the display, alarm relays, and

Teledyne Analytical Instruments

9

Image 17
Contents Teledyne Analytical Instruments Trace Oxygen AnalyzerCopyright 2009 Teledyne Instruments/ Analytical Instruments Model 3190ZSafety Messages Model 3190Z Table of Contents Maintenance Combustible GAS Usage Viii Introduction Main Features of the AnalyzerOverview Trace Oxygen Analyzer IntroductionSpan Front Panel DescriptionIntroduction Model 3190Z Measurement Parameters Readout Rear Panel DescriptionMeter Readout Rear Panel AC and DC versions ƒ Power ConnectionAlarm Page Introduction Operational TheoryZirconium Oxide Sensors Trace Oxygen Analyzer Operational TheoryOperational Theory Model 3190Z ElectronicsEffect of Pressure GeneralBlock Diagram of the Signal Processing Electronics Signal ProcessingOperational Theory Model 3190Z Unpacking the Analyzer InstallationTrace Oxygen Analyzer Installation Control Unit Installation Installation Model 3190ZExternal Probe Installation Location and MountingPower switch is located below the fuse receptacle Rear Panel Electrical Connectors for AC and DC Units Contact ID for Failsafe Relay Operation Gas Connections Installation ChecklistTrace Oxygen Analyzer Operation OperationOperation Model 3190Z Using the Function and Data Entry ButtonsSetting the Analysis Ranges HI Alarm Setting the Alarm SetpointsLO Alarm HI RangeProcedure Sensor Fail AlarmSelecting a Fixed Range or Autoranging CalibrationLED Display on StartUp Page Trace Oxygen Analyzer Maintenance MaintenanceAC Powered Units Replacing the FuseMaintenance Model 3190Z Sensor Installation or ReplacementDC Powered Units When to Replace a SensorAppendix SpecificationsTrace Oxygen Analyzer Appendix Range ID 0-10 VDCAppendix Model 3190Z Spare Parts ListMiscellaneous Drawing List