Teledyne 3190Z Selecting a Fixed Range or Autoranging, Calibration, Sensor Fail Alarm, Procedure

Page 28

Operation

Model 3190Z

 

 

 

1.Press the SET LO ALARM function button once.

2.Within 5 seconds press either the Δ or button to raise or lower the displayed value as required, until the display reads the desired concentration.

4.4.3Sensor Fail Alarm

Not active in this product.

4.5 Selecting a Fixed Range or Autoranging

The Model 3290 can operate in fixed high, fixed low, or auto- ranging mode. To change modes:

1.Press and then release the SET HI RANGE and the SET LO RANGE buttons simultaneously.

2.Within 5 seconds press either the Δ or button until Auto, Lo, or Hi displays on the LCD, as desired.

After about three seconds, the analyzer resumes monitoring in the selected range mode.

4.6 Calibration

Power up the analyzer and allow the LED reading to stabilize. Wait at least ten minutes after powering, giving the heater and bias voltages enough time to reach operating voltage. Set the alarm set-points and the full-scale ranges to the desired values.

Procedure:

1.Apply calibration gas of a known concentration. A span gas containing an oxygen content of 60-100% of the range of the analyzer is recommended.

Note: If the analyzer output goes above the high alarm set-point or below the low alarm set-point, the front panel ALARM Indicator located beneath the SET function buttons will blink. When the SPAN key is pressed, the alarm indicator will stop blinking.

2.Press the SPAN button once.

Teledyne Analytical Instruments

20

Image 28
Contents Trace Oxygen Analyzer Teledyne Analytical InstrumentsModel 3190Z Copyright 2009 Teledyne Instruments/ Analytical InstrumentsSafety Messages Model 3190Z Table of Contents Maintenance Combustible GAS Usage Viii Main Features of the Analyzer IntroductionOverview Trace Oxygen Analyzer IntroductionIntroduction Model 3190Z Front Panel DescriptionSpan Meter Readout Rear Panel DescriptionMeasurement Parameters Readout ƒ Power Connection Rear Panel AC and DC versionsAlarm Page Operational Theory IntroductionZirconium Oxide Sensors Trace Oxygen Analyzer Operational TheoryElectronics Operational Theory Model 3190ZEffect of Pressure GeneralSignal Processing Block Diagram of the Signal Processing ElectronicsOperational Theory Model 3190Z Trace Oxygen Analyzer Installation InstallationUnpacking the Analyzer Installation Model 3190Z Control Unit InstallationExternal Probe Installation Location and MountingPower switch is located below the fuse receptacle Rear Panel Electrical Connectors for AC and DC Units Contact ID for Failsafe Relay Operation Installation Checklist Gas ConnectionsOperation Trace Oxygen Analyzer OperationSetting the Analysis Ranges Using the Function and Data Entry ButtonsOperation Model 3190Z Setting the Alarm Setpoints HI AlarmLO Alarm HI RangeSensor Fail Alarm ProcedureSelecting a Fixed Range or Autoranging CalibrationLED Display on StartUp Page Maintenance Trace Oxygen Analyzer MaintenanceAC Powered Units Replacing the FuseSensor Installation or Replacement Maintenance Model 3190ZDC Powered Units When to Replace a SensorSpecifications AppendixTrace Oxygen Analyzer Appendix Range ID 0-10 VDCSpare Parts List Appendix Model 3190ZDrawing List Miscellaneous