How to Measure Blood Pressure (in AVG. Mode)

1.Wrap the arm cuff over the patient’s arm.

2.Push the ON/OFF (power) Button to turn on the power.

ON/OFF

START

HIDE

 

 

 

DEFL ATION

P-SET

 

AVG./1s

t/2nd/3rd

MODE

 

140

180

SINGLE

 

 

AVG.

MANU.

 

100

220

CHECK

 

STOP

 

260

AUTO

280

3. Set the MODE Selector to

“AVG.”.

 

ON/OFF

 

 

 

HIDE

START

 

 

 

 

DEFL ATION

P-SET

MODE

AVG./1s t/2nd/3rd

 

140 180

SINGLE

MANU.

 

AVG.

100

220

CHECK

 

260

STOP

 

 

AUTO

280

 

4.Set the P-SET (pressure set- ting) Volume to “AUTO” or the target in ation value.

When setting the P-SET to “AUTO”, turn the Volume coun- terclockwise as far as it goes until you can hear the click sound.

When the systolic pressure is expected to exceed 220 mmHg, proper in ation may not be pos- sible in “AUTO” setting.

Set the target in ation value to 30 to 40 mmHg higher than the expected systolic pressure.

ON/OFF

START

HIDE

 

 

 

DEFL ATION

P-SET

 

AVG./1s

t/2nd/3rd

MODE

 

140

180

SINGLE

 

 

AVG.

MANU.

 

100

220

CHECK

 

STOP

 

260

AUTO

280

5.Push the START Button to start the measurement.

Set the unit to the start time of the rst measurement, then the unit starts measurement auto- matically.

After displaying the results of rst measurement, the unit will measure blood pressure in the set number of times automati- cally by taking an interval.

For the setting of the number of measurements, the waiting time before start of measurement, and the interval time, refer to Page 7.

ON/OFF

START

HIDE

 

 

 

DEFL ATION

P-SET

 

AVG./1s t/2nd/3rd

MODE

140

180

SINGLE

 

AVG.

MANU.

100

220

CHECK

STOP

 

260

AUTO

280

6.Measured results are displayed.

After the measurement is completed, average values are displayed.

Each time a DEFLATION (de ation control) Button is pushed, the measurement results for each time are displayed.

SYS

mmHg

DIA

mmHg

PULSE

BPM

HIDE

7. Push the ON/OFF (power)

Button to turn off the power.

 

ON/OFF

 

HIDE

START

 

 

 

 

DEFL ATION

P-SET

MODE

AVG./1s t/2nd/3rd

 

140 180

SINGLE

MANU.

 

AVG.

100

220

CHECK

 

260

STOP

 

 

AUTO

280

 

To stop the measurement during the measurement:

Push the STOP Button.

To start the stopped measurement again:

Push the START Button.

17

Image 17
Contents HEM-907 Table of ContentsIntended Use Exemptions DC8V Names and Functions of the PartsExample of display Number of irregular pulse wavesManual de ation control function Pressure setting function P-SETNon-display function Hide Procedure to change the set values Setting mode change functionList of Functions for each Mode Items to set Set valueExamples of signs ContentsWhen an error is generated or the measured value is doubtful Requests from Omron Measures to take at the time of trouble or accident MaintenanceDurability Options Components of the ProductPrepare the patient to be ready to wrap the arm cuff Securely connect the air tubeAdjust the level of the arm cuff to the level of the heart How to Apply the Arm CuffList of Measurement Modes How to Measure Blood PressureSet the Mode Selector to How to Measure Blood Pressure in Single ModePush the Start Button to start the measurement Measured results are dis- playedPush the Start Button Push the Stop ButtonWrap the arm cuff over the patient’s arm Measured results are displayedHow to Measure Blood Pressure in MANU. Mode Check result How to Measure Blood Pressure in Check ModeHow to use the blood pressure monitor Battery life Installation and Replacement of Battery PackBattery low Charging timeHow to Clean the Unit after Use Error code Cause How to correct List of Error CodesProblem What to inspect How to correct TroubleshootingHEM-907-E / HEM-907-E7 Speci cationsOmron Healthcare Europe B.V EMC & Weee