Omron Healthcare HEM-907 instruction manual How to Measure Blood Pressure in MANU. Mode

Page 18

How to Measure Blood Pressure (in MANU. Mode)

1.Wrap the arm cuff over the patient’s arm.

3.Push the ON/OFF (power) Button to turn on the power.

ON/OFF

START

HIDE

 

 

 

DEFL ATION

P-SET

 

AVG./1s t/2nd/3rd

MODE

140

180

SINGLE

 

AVG.

MANU.

100

220

CHECK

STOP

 

260

AUTO

280

2.Place the stethoscope on the patient’s arm.

4.Set the MODE Selector to “MANU.”.

ON/OFF

START

HIDE

 

 

DEFL ATION

P-SET

AVG./1s

t/2nd/3rd

MODE

 

140

180

SINGLE

 

 

AVG.

MANU.

 

100

220

CHECK

 

STOP

 

260

AUTO

280

5.Set the P-SET (pressure set- ting) Volume to “AUTO” or the target in ation value.

When setting the P-SET to “AUTO”, turn the Volume coun- terclockwise as far as it goes un- til you can hear the click sound.

When the systolic pressure is expected to exceed 220 mmHg, proper in ation may not be pos- sible in “AUTO” setting.

Set the target in ation value to 30 to 40 mmHg higher than the expected systolic pressure.

ON/OFF

START

HIDE

 

 

DEFL ATION

P-SET

AVG./1s t/2nd/3rd

MODE

140

180

SINGLE

 

AVG.

MANU.

100

220

CHECK

STOP

 

260

AUTO

280

6.Push the START Button to start the measurement.

In the “AUTO” setting, the unit stops in ation at the pres- sure considered to be 30 to

40 mmHg above the expected systolic pressure, then start de ation.

The unit can be re-in ated

only while the START Button is pushed.

The pressure display is syn- chronized with the pulse rate and renewed.

Each time the DEFLATION (de ation control) Button is pushed, the unit de ates by 5~10 mmHg.

When you press the STOP But- ton, the unit de ates fast and ends the operation.

ON/OFF

START

HIDE

 

 

 

DEFL ATION

P-SET

 

AVG./1s t/

2nd/3rd

MODE

 

140

180

SINGLE

 

 

AVG.

MANU.

 

100

220

CHECK

 

STOP

 

260

AUTO

280

7. Push the ON/OFF (power)

Button to turn off the power.

 

ON/OFF

 

 

 

HIDE

START

 

 

 

 

DEFL ATION

P-SET

MODE

AVG./1s t/2nd/3rd

 

140 180

SINGLE

MANU.

 

AVG.

100

220

CHECK

 

260

STOP

 

 

AUTO

280

 

18

Image 18
Contents Table of Contents HEM-907Intended Use Exemptions Names and Functions of the Parts DC8VNumber of irregular pulse waves Example of displayPressure setting function P-SET Non-display function HideManual de ation control function List of Functions for each Mode Setting mode change functionProcedure to change the set values Items to set Set valueContents Examples of signsWhen an error is generated or the measured value is doubtful Requests from Omron Maintenance DurabilityMeasures to take at the time of trouble or accident Components of the Product OptionsSecurely connect the air tube Prepare the patient to be ready to wrap the arm cuffHow to Apply the Arm Cuff Adjust the level of the arm cuff to the level of the heartHow to Measure Blood Pressure List of Measurement ModesPush the Start Button to start the measurement How to Measure Blood Pressure in Single ModeSet the Mode Selector to Measured results are dis- playedWrap the arm cuff over the patient’s arm Push the Stop ButtonPush the Start Button Measured results are displayedHow to Measure Blood Pressure in MANU. Mode How to Measure Blood Pressure in Check Mode How to use the blood pressure monitorCheck result Battery low Installation and Replacement of Battery PackBattery life Charging timeHow to Clean the Unit after Use List of Error Codes Error code Cause How to correctTroubleshooting Problem What to inspect How to correctSpeci cations HEM-907-E / HEM-907-E7Omron Healthcare Europe B.V EMC & Weee