Omron Healthcare HEM-907 How to Measure Blood Pressure in Check Mode, Check result

Page 19

How to Measure Blood Pressure (in CHECK Mode)

Display of pressure can be con rmed by the CHECK Mode (yearly procedure).

What you need to prepare

(1)Well-adjusted reference pressure monitor (including rubber ball),

(2)T-shaped tube,

(3)Two rubber tubes, and (4) something in the cylindrical shape to wrap the arm cuff such as the tea container (sturdy one that will not break or deform with pressure)

*Measurement error may occur even with the mercury blood pressure monitor because of insuf cient amount of mercury in the mercury, dirty glass tube, or clogged glass tube cap.

How to use the blood pressure monitor

1.Connect the mercury blood pressure monitor, rubber ball, and this unit with the T-shaped tube as shown in the Figure.

2.Push the ON/OFF (power) Button to turn on the power.

3.Set the MODE Selector to “CHECK”.

4.Tightly wrap the arm cuff over a sturdy cylindrical object.

5.Close the air release valve of rubber ball and in ate the monitor to a certain value (pressure to be checked).

 

 

 

SYS

 

 

 

 

 

 

mmHg

 

 

 

DIA

 

 

 

 

P

ULSE

mmHg

 

 

 

 

 

 

 

 

 

B

 

 

 

 

 

PM AC

 

 

 

 

 

HIDE

 

P-

 

 

 

 

140

SET

 

 

 

 

180

 

MODE

100

 

 

 

 

 

 

 

 

 

 

220

AVG.

SINGLE

 

 

 

 

 

MANU.

AUTO

 

 

260

 

CHECK

 

 

 

280

 

 

 

ON/OFF

 

ST

 

ART

A

DEF

VG./1 LATION

 

st /2nd/3rd

 

STOP

6.Compare the value displayed on this monitor and that on the mercury.

7.Open the air release valve of rubber ball to de ate.

CHECK result

(1)In the measuring rule, the test tolerance of blood pressure monitor (accuracy of pressure when shipped from the factory) should not exceed 3 mmHg.

(2)If the CHECK result is +/- 3 mmHg or above, contact the dealer where you purchased this unit or the nearest OMRON dealer.

19

Image 19
Contents HEM-907 Table of ContentsIntended Use Exemptions DC8V Names and Functions of the PartsExample of display Number of irregular pulse wavesNon-display function Hide Pressure setting function P-SETManual de ation control function Items to set Set value Setting mode change functionProcedure to change the set values List of Functions for each ModeExamples of signs ContentsWhen an error is generated or the measured value is doubtful Requests from Omron Durability MaintenanceMeasures to take at the time of trouble or accident Options Components of the ProductPrepare the patient to be ready to wrap the arm cuff Securely connect the air tubeAdjust the level of the arm cuff to the level of the heart How to Apply the Arm CuffList of Measurement Modes How to Measure Blood PressureMeasured results are dis- played How to Measure Blood Pressure in Single ModeSet the Mode Selector to Push the Start Button to start the measurementMeasured results are displayed Push the Stop ButtonPush the Start Button Wrap the arm cuff over the patient’s armHow to Measure Blood Pressure in MANU. Mode How to use the blood pressure monitor How to Measure Blood Pressure in Check ModeCheck result Charging time Installation and Replacement of Battery PackBattery life Battery lowHow to Clean the Unit after Use Error code Cause How to correct List of Error CodesProblem What to inspect How to correct TroubleshootingHEM-907-E / HEM-907-E7 Speci cationsOmron Healthcare Europe B.V EMC & Weee