Names and Functions of the Parts
Explanation of Functions
(1) Pressure setting function P-SET
In ation can be set by AUTO (automatic setting) or the target value.
AUTO (automatic setting): In the SINGLE (single measurement), AVG. (average), and MANU. (auscultation) Modes, the monitor estimates the systolic blood pressure value during in ation and automatically in ates to the proper value.
Target value setting: The monitor in ates to the target value.Pressure value is set to 30 to 40 mmHg above the expected systolic pressure.
(2) Non-display function HIDE
A function not to display the results of measured blood pressures. However, the pressure values during the measurement are displayed.
This function can be used when SINGLE (single measurement) or AVG. (average) Mode is selected.
(3) Manual de ation control function
A function to accelerate the de ation speed by pushing a DEFLATION Button during de ation.
This function can be used when MANU. (auscultation) Mode is selected.