Troubleshooting

If a trouble occurs while using the unit, please check the following.

 

 

 

Problem

What to inspect

How to correct

 

Is the arm cuff wrapped correctly?

Wrap the arm cuff correctly, and measure

 

again.

 

 

 

 

 

 

Is the patient moving his/her arm or body

Ask the patient not to move his/her arm or

 

body during measurement, and measure

The unit in ates to abnormally

during in ation?

again.

high (low) pressure.

 

 

 

 

Does this trouble occur with a speci c

 

 

patient?

Check the measurement by the stetho-

 

For patients having arrhythmia, the unit

scope.

 

may not in ate properly

 

 

 

 

 

 

After checking the patient with the stetho-

 

First of all, check the patient’s condition.

scope as needed, take a proper mea-

 

sure in accordance with the “list of error

 

 

 

 

codes”.

 

 

 

 

Is the patient moving his/her arm or body

Ask the patient not to move his/her arm or

 

body during measurement, and measure

The unit cannot measure

during measurement?

again.

 

blood pressure.

 

 

Does the patient have an arrhythmia?

Check the patient with a stethoscope.

The blood pressure values are

extremely high (low).

 

 

Is the size of the arm cuff and the wrapping

Select the arm cuff in accordance to the

 

 

of arm cuff correct?

circumference of the patient’s arm, wrap it

 

correctly, then measure again.

 

 

 

 

 

 

Is the level of the brachium to which the

Keep the level of the brachium to which

 

arm cuff is wrapped at the same level as

the arm cuff is wrapped at the same level

 

the heart?

as the heart, then measure again.

 

 

 

23

Image 23
Contents HEM-907 Table of ContentsIntended Use Exemptions DC8V Names and Functions of the PartsExample of display Number of irregular pulse wavesManual de ation control function Pressure setting function P-SETNon-display function Hide Items to set Set value Setting mode change functionProcedure to change the set values List of Functions for each ModeExamples of signs ContentsWhen an error is generated or the measured value is doubtful Requests from Omron Measures to take at the time of trouble or accident MaintenanceDurability Options Components of the ProductPrepare the patient to be ready to wrap the arm cuff Securely connect the air tubeAdjust the level of the arm cuff to the level of the heart How to Apply the Arm CuffList of Measurement Modes How to Measure Blood PressureMeasured results are dis- played How to Measure Blood Pressure in Single ModeSet the Mode Selector to Push the Start Button to start the measurementMeasured results are displayed Push the Stop ButtonPush the Start Button Wrap the arm cuff over the patient’s armHow to Measure Blood Pressure in MANU. Mode Check result How to Measure Blood Pressure in Check ModeHow to use the blood pressure monitor Charging time Installation and Replacement of Battery PackBattery life Battery lowHow to Clean the Unit after Use Error code Cause How to correct List of Error CodesProblem What to inspect How to correct TroubleshootingHEM-907-E / HEM-907-E7 Speci cationsOmron Healthcare Europe B.V EMC & Weee