Omron Healthcare HEM-907 How to Measure Blood Pressure, List of Measurement Modes

Page 15

How to Measure Blood Pressure

Warning

If any abnormal matter occurs during the mea- surement such as the in ation does not stop, remove the arm cuff or pull the air tube from the main unit.

You may suffer peripheral neuropathy.

Requests from OMRON

Be sure that the patient should not touch the unit. Be sure to monitor all the time that the unit and the patient are not abnormal.

Do not use the unit in a vehicle.

Caution

In the following case, con rm the measurement with the stethoscope.

(1) When an irregular pulse wave is displayed

· An error may be displayed when an external vibra- tion is applied or the patient moves his/her body.

(2)When an error is generated or the measured value is doubtful

·An error may be displayed when an external vibra- tion is applied or the patient moves his/her body.

·An error may be generated in the measured valueor the blood pressure may not be able to measure for the patient having peripheral circulatory disturbance or extreme hypotension.

·An error may be generated in the measured valueor the blood pressure may not be able to measure for the patient having short variation of blood pressure such as arrhythmia.

Do not use a portable phone near the unit. The unit may malfunction.

List of Measurement Modes

To measure

only once

To measure three

times

(or two times) consecutively

To measure

by using a

stethoscope

To con rm the

accuracy of

pressure display

SINGLE Mode

AVG. Mode

MANU. Mode

CHECK Mode

Refer to Page

16.

Refer to Page

17.

Refer to Page

18.

Refer to Page

19.

15

Image 15
Contents HEM-907 Table of ContentsIntended Use Exemptions DC8V Names and Functions of the PartsExample of display Number of irregular pulse wavesPressure setting function P-SET Non-display function HideManual de ation control function Items to set Set value Setting mode change functionProcedure to change the set values List of Functions for each ModeExamples of signs ContentsWhen an error is generated or the measured value is doubtful Requests from Omron Maintenance DurabilityMeasures to take at the time of trouble or accident Options Components of the ProductPrepare the patient to be ready to wrap the arm cuff Securely connect the air tubeAdjust the level of the arm cuff to the level of the heart How to Apply the Arm CuffList of Measurement Modes How to Measure Blood PressureMeasured results are dis- played How to Measure Blood Pressure in Single ModeSet the Mode Selector to Push the Start Button to start the measurementMeasured results are displayed Push the Stop ButtonPush the Start Button Wrap the arm cuff over the patient’s armHow to Measure Blood Pressure in MANU. Mode How to Measure Blood Pressure in Check Mode How to use the blood pressure monitorCheck result Charging time Installation and Replacement of Battery PackBattery life Battery lowHow to Clean the Unit after Use Error code Cause How to correct List of Error CodesProblem What to inspect How to correct TroubleshootingHEM-907-E / HEM-907-E7 Speci cationsOmron Healthcare Europe B.V EMC & Weee