TROUBLESHOOTING
If any abnormality arises during use, please check the following points
SYMPTOMS | POSSIBLE | CORRECTION | |
CAUSES | |||
|
| ||
|
|
| |
Unit does not turn | Batteries have | Replace them with | |
on when the START/ | run down. | two new AA alkaline | |
STOP |
| batteries. | |
button is pushed. |
|
| |
Battery polarities | |||
| have been positioned | in the correct | |
| incorrectly. | positions. | |
|
|
| |
EE measurement | Cuff has been placed | Wrap the cuff properly | |
error symbol shown | incorrectly. | so that it is positioned | |
on display or the |
| correctly. | |
blood pressure |
|
| |
Did you talk or move | Measure again. | ||
value is displayed | during measurement? | Keep arm steady | |
excessively low |
| during measurement. | |
Shaking of the arm | |||
(or high). | with the cuff on. |
| |
|
| ||
|
|
| |
E1 error symbol | Air circuit abnormality. | Check cuff connection. | |
shown on display | Cuff tube may not be | Measure again. | |
| plugged into monitor |
| |
| correctly. |
| |
|
|
| |
E2 error symbol | Inflation pressure | Turn the unit off, then | |
shown on display | exceeding 300 mmHg. | measure again. | |
|
|
| |
E3 error symbol | Error determining | Measure again. | |
shown on display | measurement data. |
| |
|
|
|
Note: If the unit still does not work, contact HoMedics Consumer Relations. Under no circumstance should you disassemble or attempt to repair the unit by yourself. Contact information for HoMedics Consumer Relations Department can be found on the warranty page.
27
SPECIFICATIONS
Power Source: | Four 1.5V DC (AA) batteries | |
|
| |
Measurement | Oscillometric | |
Method: | ||
| ||
Measurement Range: | Pressure: 40~280 millimeters mercury (mmHg) | |
Pulse: 40~199 beats/minute | ||
Accuracy: | Pressure: ±3 mmHg | |
Pulse: ±5% of reading | ||
| ||
Pressure Sensor: | ||
|
| |
Inflation: | Manual Inflation | |
Deflation: | Automatic Pressure release valve | |
Memory Capacity: | 60 memories | |
1 minute after last button operation | ||
Operation | Temperature: 50°F~104°F (10°C ~40°C) | |
Environment: | Humidity: 15% – 90% R.H. | |
Storage Environment: | Temperature: | |
Humidity: Less than 90% R.H. | ||
Weight: | .43 lbs/193 grams (without batteries) | |
Arm Circumference: | Standard size adult arm cuff: | |
Dimensions: | 5”(L) x 4.1”(W) x 2.7”(H) | |
126mm(L) x 105mm(W) x 68mm(H) | ||
| ||
| Internally Powered equipment | |
| Type BF Equipment | |
| ||
| ingress of water | |
| Not suitable for use in presence of flammable | |
| anesthetic mixture with air, oxygen or | |
| nitrous oxide. | |
| Continuous operation with | |
|
|
Note: These specifications are subject to change without notice.
28
5/01/10 1:27 PM