TROUBLESHOOTING
If any abnormality arises during use, please check the following points
SYMPTOMS | POSSIBLE | CORRECTION | |
CAUSES | |||
|
| ||
|
|
| |
Unit does not turn on | Batteries have | Replace them with | |
when the | run down. | two new AA alkaline | |
START/STOP |
| batteries. | |
button is pushed. |
|
| |
Battery polarities | |||
| have been positioned | in the correct | |
| incorrectly. | positions. | |
|
|
| |
EE measurement | Cuff has been placed | Wrap the cuff | |
error symbol shown | incorrectly. | properly so that it is | |
on display or the |
| positioned correctly. | |
blood pressure value |
|
| |
Did you talk or move | Measure again. | ||
is displayed | during measurement? | Keep arm steady | |
excessively low |
| during measurement. | |
Shaking of the arm | |||
(or high). | with the cuff on. |
| |
|
| ||
|
|
| |
E1 error symbol | Air circuit abnormality. | Check cuff connection. | |
shown on display | Cuff tube may not be | Measure again. | |
| plugged into monitor |
| |
| correctly. |
| |
|
|
| |
E2 error symbol | Inflation pressure | Turn the unit off, then | |
shown on display | exceeding 300 mmHg. | measure again. | |
|
|
| |
E3 error symbol | Error determining | Measure again. | |
shown on display | measurement data. |
| |
|
|
| |
CA appears on | Recalibration alert. | Refer to page 25 for | |
display | more information. | ||
| |||
|
|
|
Note: If the unit still does not work, contact HoMedics Consumer Relations. Under no circumstance should you disassemble or attempt to repair the unit by yourself. Contact information for HoMedics Consumer Relations Department can be found on the warranty page.
SPECIFICATIONS
Power Source: | Four 1.5V DC (AA) batteries | |
|
| |
Measurement | Oscillometric | |
Method: | ||
| ||
|
| |
Measurement | Pressure: 40~280 millimeters mercury (mmHg) | |
Range: | Pulse: 40~199 beats/minute | |
|
| |
Accuracy: | Pressure: ±3 mmHg | |
Pulse: ±5% of reading | ||
| ||
|
| |
Pressure Sensor: | ||
|
| |
Inflation: | Manual Inflation | |
|
| |
Deflation: | Automatic Pressure release valve | |
Memory Capacity: | 60 memories | |
1 minute after last button operation | ||
|
| |
Operation | Temperature: 50°F~104°F (10°C ~40°C) | |
Environment: | Humidity: Less than 85% R.H. | |
|
| |
Storage | Temperature: | |
Environment: | Humidity: Less than 85% R.H. | |
Weight: | .40 lbs/181 grams (without batteries) | |
|
| |
Arm | Standard size adult arm cuff: | |
Circumference: | ||
|
| |
Dimensions: | 4.9"(L) x 4"(W) x 2.5"(H) | |
125mm(L) x 102mm(W) x 63mm(H) | ||
| ||
|
| |
| Type BF | |
| Device and cuff are designed to provide | |
| special protection against electrical shocks. | |
|
|
Note: These specifications are subject to change without notice.