Omega Engineering CYD201 and CYD208 manual Latched And Unlatched Alarms, Softcal Compensations

Page 25

OMEGA Model CYD201/CYD208 User’s Manual

2.Hold UNITS until the desired temperature displays. UNITS is a toggle; if it is released and pressed again, the temperature direction reverses. If the temperature display increases, the alarm is a high setpoint. If the temperature display decreases, the alarm is a low setpoint.

3.Release UNITS when the desired setpoint displays. For a Model CYD208, press CHANNEL to display the desired channel in the channel window and repeat the steps above to set the alarm for each channel

4.Move the SET switch back to position 0 to enable the alarm. When it triggers, an alarm status light appears in the upper left of the temperature display.

The alarm can be connected to another device which triggers when the alarm activates. The 3-contact terminal block is present on the rear panel as J3 ALARM. The alarm contacts are designated 1 COM 2 with 1 representing the normally open state and 2 representing the normally closed state.

3.5.2Latched And Unlatched Alarms

Alarms are either latched or unlatched. The alarm is latched when the LATCH switch is in position 1: the alarm turns On when triggered by the alarm setpoint, but will not automatically turn Off when the temperature returns to within the high and low setpoint range. The alarm is unlatched when the LATCH switch is in position 0; the alarm turns On when triggered by the alarm setpoint, and automatically turns Off when temperatures return to within the high and low setpoint range.

3.5.3Alarm Fix Function (Model CYD208 Only)

Set the FIX switch on the Model CYD208 rear panel to OFF (position 0) to continuously update the alarm relay, depending on the alarm setpoint and sensor temperature. If the FIX switch is ON (position 1), the relay updates only when channel 1 input is active.

3.6SOFTCAL™ COMPENSATIONS

SoftCal™ is a simple, instrument-configured software calibration that improves system accuracy over a specified temperature range. It reduces the error between a CY-7 diode and the Standard Curve 10 used by the instrument. In short, SoftCal™ generates inexpensive calibrations for CY-7 sensors used with OMEGA temperature controllers and monitors.

Operation

3-3

Image 25
Contents M789-038A June Return Requests / Inquiries WarrantyTable of Contents Table of Contents List of Illustrations This Page Intentionally Left Blank Table of Contents General Chapter IntroductionModel CYD201/CYD208 General Description Resolution Model CYD201/CYD208 SpecificationsHandling Cryogenic Storage Dewars LHe and LN2 Safety PrecautionsHandling Liquid Helium and Liquid Nitrogen Electrostatic Discharge Recommended First AidIdentifying Esds Components Safety SummaryHandling Esds Components Ground The InstrumentDo Not Operate In An Explosive Atmosphere Safety SymbolsKeep Away From Live Circuits Do Not Substitute Parts Or Modify InstrumentInspection and Unpacking Chapter InstallationRepackaging for Shipment Sensor Installation Recommendations Power and Ground RequirementsTwo-Lead Measurement Two-Lead Vs Four-Lead MeasurementsFour-Lead Measurement Sensor Mounting Connecting Leads To The SensorMeasurement Errors Due To AC Noise J1 Input Sensor Input ConnectionsTerminal Description Rack Mounting Sensor Curve DefinitionModel CYD208-DIN Rack Mounting Hole SizesPower UP Errors Initial Power UP SequenceThis Page Intentionally Left Blank Installation Units KEY Chapter OperationScan Mode Model CYD208 Only Alarm SetpointSetting Dwell Times Model CYD208 Only Alarm OperationAlarm Fix Function Model CYD208 Only Latched And Unlatched AlarmsSoftcal Compensations SoftCal Calibration Procedure Erasing SoftCal Compensations Verifying SoftCal OperationOperation Serial Interface Connections Chapter Remote OperationSerial Interface J2 Serial I/O Serial I/O RJ-11 Connector Pin DefinitionsSample Basic Program Serial Interface OperationSample Quick Basic 4.0 Program Serial Interface Command SummaryInput ReturnedInput Hxxx.x Returned Nothing Input F0x Returned NothingInput Lxxx.x Returned Nothing Input R Returned NothingHigh or Low Parameter H = high alarm, L = low alarm Switch ID and Alarm DataInput YCx Returned Nothing End Scanning Model CYD208 OnlyCurrent channel,sign,sensor reading,units, alarm statusCRLF Sign,sensor reading,units,alarm status CrlfScan Status Model CYD208 Only This Page Intentionally Left Blank Remote Operation Model CYD201 Rear Panel Connections Chapter ServiceModel CYD208 Rear Panel Connections Model CYD208 Rear Panel ConnectionsGeneral Maintenance Error Code TroubleshootingLine Voltage Configuration Fuse ReplacementRecalibration Current Source CalibrationModel CYD201 Model CYD208 Serial Interface Cable and Adapters 2 A/D Converter CalibrationModel CYD200-B RJ-11 to DE-9 Adapter Wiring Details Page Accessories Model Description of Accessory Chapter Options and AccessoriesModel CYD201/CYD208 Wires Lsci P/N Description of Cable Model Description of AccessoryDIN Series DT-420 Model CYD201/CYD208 Sensors Sensor no Description of SensorSeries DT-450 Series CY-7Options and Accessories A1.0 General Appendix a Curve TablesTable A-2. Curve 1 DT-500DI-8A Voltage-Temp. Characteristics Table A-3. Curve 2 DT-500DRC-D Voltage-Temp. Characteristics 6591 18.0 Table A-5. Curve 4 CTI Diode Voltage-Temp. Characteristics Table A-6. Curve 5 DT-500DI-8C Voltage-Temp. Characteristics Table A-7. Curve 6 CY-7 Voltage-Temp. Characteristics 305.0 50689 Page Page PRESSURE/STRAIN TemperatureFlow