Omega Engineering CYD201 and CYD208 manual SoftCal Calibration Procedure

Page 26

OMEGA Model CYD201/CYD208 User’s Manual

SoftCal™ calibrations are made at three temperature points: liquid helium (4.2 K), liquid nitrogen (about 77 K), and 305 K. Below is the accuracy* of the CY-7-SD-13 sensor:

+0.5 K

2 K to <30 K

+0.25 K

345

K to <375 K

+0.25 K

30

K to <60 K

+1.0 K

375

K to 475 K

+0.15 K

60

K to <345 K

 

 

 

*These values generally apply to all silicon diode sensors. Only two-point SoftCal™ calibrations appropriate for CY-7 series Band 11, 11A sensors.

This section covers the SoftCal™ Calibration Procedure (Paragraph 3.6.1), Verifying SoftCal™ Operation (Paragraph 3.6.2), and Erasing SoftCal™ Compensations (Paragraph 3.6.3).

3.6.1SoftCal™ Calibration Procedure

1.Turn on thermometer 30 minutes prior to operation.

2.Place the SET switch in position 1.

3.Hold UNITS until 0 kelvin (or equivalent in °C or °F) displays. The key acts as a toggle. If the display rises, release and press again.

4.Press CAL ENABLE on rear panel with a pen tip. The display reads "-SOF-" to indicate the unit is ready to erase the current SoftCal™ calibration. Within 2 seconds press UNITS again. The temperature display changes from "-SOF-" to the current SoftCal™ setting.

5.Hold UNITS until the sensor temperature displays. For example, if setting SoftCal™ for liquid helium, the display reads 4.2 K. If setting for ice point, the display reads 0 °C.

6.Verify sensor stabilization at calibration temperature.

7.Press CAL ENABLE again. "-SOF-" again displays to indicate the unit is ready to accept the calibration point. Within 2 seconds, press UNITS to enter a new calibration point. (If UNITS is not pressed within 2 seconds, the display returns to the alarm temperature). After 15 seconds, the alarm setpoint temperature displays.

8.To enter more than one point, go back to step 5. NOTE: One point may be entered between 1.4 K to 9.9 K. No point may be entered between 10 K and 40 K. Two points may be entered above 40 K.

9.Return the SET switch to position 0.

3-4

Operation

Image 26
Contents M789-038A June Warranty Return Requests / InquiriesTable of Contents Table of Contents List of Illustrations This Page Intentionally Left Blank Table of Contents Model CYD201/CYD208 General Description Chapter IntroductionGeneral Model CYD201/CYD208 Specifications ResolutionHandling Liquid Helium and Liquid Nitrogen LHe and LN2 Safety PrecautionsHandling Cryogenic Storage Dewars Recommended First Aid Electrostatic DischargeHandling Esds Components Safety SummaryIdentifying Esds Components Ground The InstrumentKeep Away From Live Circuits Safety SymbolsDo Not Operate In An Explosive Atmosphere Do Not Substitute Parts Or Modify InstrumentRepackaging for Shipment Chapter InstallationInspection and Unpacking Power and Ground Requirements Sensor Installation RecommendationsFour-Lead Measurement Two-Lead Vs Four-Lead MeasurementsTwo-Lead Measurement Connecting Leads To The Sensor Sensor MountingMeasurement Errors Due To AC Noise Terminal Description Sensor Input ConnectionsJ1 Input Sensor Curve Definition Rack MountingHole Sizes Model CYD208-DIN Rack MountingInitial Power UP Sequence Power UP ErrorsThis Page Intentionally Left Blank Installation Chapter Operation Units KEYSetting Dwell Times Model CYD208 Only Alarm SetpointScan Mode Model CYD208 Only Alarm OperationSoftcal Compensations Latched And Unlatched AlarmsAlarm Fix Function Model CYD208 Only SoftCal Calibration Procedure Verifying SoftCal Operation Erasing SoftCal CompensationsOperation Serial Interface Chapter Remote OperationSerial Interface Connections Serial I/O RJ-11 Connector Pin Definitions J2 Serial I/OSerial Interface Operation Sample Basic ProgramInput Serial Interface Command SummarySample Quick Basic 4.0 Program ReturnedInput Lxxx.x Returned Nothing Input F0x Returned NothingInput Hxxx.x Returned Nothing Input R Returned NothingInput YCx Returned Nothing Switch ID and Alarm DataHigh or Low Parameter H = high alarm, L = low alarm End Scanning Model CYD208 OnlyScan Status Model CYD208 Only Sign,sensor reading,units,alarm status CrlfCurrent channel,sign,sensor reading,units, alarm statusCRLF This Page Intentionally Left Blank Remote Operation Chapter Service Model CYD201 Rear Panel ConnectionsModel CYD208 Rear Panel Connections Model CYD208 Rear Panel ConnectionsError Code Troubleshooting General MaintenanceFuse Replacement Line Voltage ConfigurationModel CYD201 Model CYD208 Current Source CalibrationRecalibration 2 A/D Converter Calibration Serial Interface Cable and AdaptersModel CYD200-B RJ-11 to DE-9 Adapter Wiring Details Page Chapter Options and Accessories Accessories Model Description of AccessoryDIN Model Description of AccessoryModel CYD201/CYD208 Wires Lsci P/N Description of Cable Series DT-450 Model CYD201/CYD208 Sensors Sensor no Description of SensorSeries DT-420 Series CY-7Options and Accessories Appendix a Curve Tables A1.0 GeneralTable A-2. Curve 1 DT-500DI-8A Voltage-Temp. Characteristics Table A-3. Curve 2 DT-500DRC-D Voltage-Temp. Characteristics 6591 18.0 Table A-5. Curve 4 CTI Diode Voltage-Temp. Characteristics Table A-6. Curve 5 DT-500DI-8C Voltage-Temp. Characteristics Table A-7. Curve 6 CY-7 Voltage-Temp. Characteristics 305.0 50689 Page Page Flow TemperaturePRESSURE/STRAIN