OMEGA Model CYD201/CYD208 User’s Manual
5.7.2A/D Converter Calibration
NOTE: Current source calibration must be performed before A/D converter calibration.
1.Allow 30 minute
2.Perform current source calibration.
3.Configure 125 kresistor as shown in Figure
4.Wait 10 seconds for resistor voltage to settle.
5.Press CAL ENABLE, then within 2 seconds press UNITS. The display window shows
6.Repeat procedure with 25 kresistor.
7.For Model CYD208, repeat procedure for other 7 channels. See Figure
5.8SERIAL INTERFACE CABLE AND ADAPTERS
6 |
|
|
5 | TxD | YELLOW |
4 | Gnd | GREEN |
3 | Gnd | RED |
2 | RxD | BLACK |
1 |
|
|
6 5 4 3 2 1
Figure 5-5. Model CYD200-J10 RJ-11 Cable Assembly Wiring Details