5.3ERROR CODE TROUBLESHOOTING

On power up, the CYD201/CYD208 checks internal memory. If a problem exists, an error message displays on the front panel of the instrument.

Er01 indicates a hardware problem in the instrument memory. This error is not user-correctable.

Er02 indicates a soft error in the instrument memory. To correct this error, use the following procedure.

1.Power up the unit and allow it to display Er02.

2.Close DIP Switch 1 (top of the switch pressed in). Leave the switch closed for at least 5 seconds, then open DIP switch 1 (bottom of the switch pressed in).

3.Verify the Model CYD201/CYD208 display goes through a normal power up sequence and then displays 499.9 K.

4.The input(s) of the Model CYD201/CYD208 must now be recalibrated per in procedure in Paragraph 5.7 before the unit can be used.

OL indicates a voltage input overload. This can be caused by an open sensor or diode sensor wired backwards.

Before calling the factory about a persistent problem, try the procedure below:

WARNING: This procedure erases calibration constants stored in Non- Volatile RAM. If this procedure is used, recalibrate the instrument.

1.With power turned Off, press and hold CAL ENABLE on the back panel. While holding CAL ENABLE, turn instrument power On.

2.If the Model CYD201/CYD208 displays Er02, follow the Calibration procedure in Paragraph 5.7. If Er01 still displays or if the Model CYD201/CYD208 does not respond, contact OMEGA Service.

5.4GENERAL MAINTENANCE

Clean the CYD201/CYD208 periodically to remove dust, grease and other contaminants. Clean the front and back panels and case with a soft cloth dampened with a mild detergent and water solution.

NOTE: Do not use aromatic hydrocarbons or chlorinated solvents to clean the Model CYD201/CYD208. They may react with the silk screen printing on the back panel.

Service

5-3

Image 39
Contents M789-038A June Return Requests / Inquiries WarrantyTable of Contents Table of Contents List of Illustrations This Page Intentionally Left Blank Table of Contents Chapter Introduction GeneralModel CYD201/CYD208 General Description Resolution Model CYD201/CYD208 SpecificationsLHe and LN2 Safety Precautions Handling Cryogenic Storage DewarsHandling Liquid Helium and Liquid Nitrogen Electrostatic Discharge Recommended First AidGround The Instrument Safety SummaryIdentifying Esds Components Handling Esds ComponentsDo Not Substitute Parts Or Modify Instrument Safety SymbolsDo Not Operate In An Explosive Atmosphere Keep Away From Live CircuitsChapter Installation Inspection and UnpackingRepackaging for Shipment Sensor Installation Recommendations Power and Ground RequirementsTwo-Lead Vs Four-Lead Measurements Two-Lead MeasurementFour-Lead Measurement Sensor Mounting Connecting Leads To The SensorMeasurement Errors Due To AC Noise Sensor Input Connections J1 InputTerminal Description Rack Mounting Sensor Curve DefinitionModel CYD208-DIN Rack Mounting Hole SizesPower UP Errors Initial Power UP SequenceThis Page Intentionally Left Blank Installation Units KEY Chapter OperationAlarm Operation Alarm SetpointScan Mode Model CYD208 Only Setting Dwell Times Model CYD208 OnlyLatched And Unlatched Alarms Alarm Fix Function Model CYD208 OnlySoftcal Compensations SoftCal Calibration Procedure Erasing SoftCal Compensations Verifying SoftCal OperationOperation Chapter Remote Operation Serial Interface ConnectionsSerial Interface J2 Serial I/O Serial I/O RJ-11 Connector Pin DefinitionsSample Basic Program Serial Interface OperationReturned Serial Interface Command SummarySample Quick Basic 4.0 Program InputInput R Returned Nothing Input F0x Returned NothingInput Hxxx.x Returned Nothing Input Lxxx.x Returned NothingEnd Scanning Model CYD208 Only Switch ID and Alarm DataHigh or Low Parameter H = high alarm, L = low alarm Input YCx Returned NothingSign,sensor reading,units,alarm status Crlf Current channel,sign,sensor reading,units, alarm statusCRLFScan Status Model CYD208 Only This Page Intentionally Left Blank Remote Operation Model CYD201 Rear Panel Connections Chapter ServiceModel CYD208 Rear Panel Connections Model CYD208 Rear Panel ConnectionsGeneral Maintenance Error Code TroubleshootingLine Voltage Configuration Fuse ReplacementCurrent Source Calibration RecalibrationModel CYD201 Model CYD208 Serial Interface Cable and Adapters 2 A/D Converter CalibrationModel CYD200-B RJ-11 to DE-9 Adapter Wiring Details Page Accessories Model Description of Accessory Chapter Options and AccessoriesModel Description of Accessory Model CYD201/CYD208 Wires Lsci P/N Description of CableDIN Series CY-7 Model CYD201/CYD208 Sensors Sensor no Description of SensorSeries DT-420 Series DT-450Options and Accessories A1.0 General Appendix a Curve TablesTable A-2. Curve 1 DT-500DI-8A Voltage-Temp. Characteristics Table A-3. Curve 2 DT-500DRC-D Voltage-Temp. Characteristics 6591 18.0 Table A-5. Curve 4 CTI Diode Voltage-Temp. Characteristics Table A-6. Curve 5 DT-500DI-8C Voltage-Temp. Characteristics Table A-7. Curve 6 CY-7 Voltage-Temp. Characteristics 305.0 50689 Page Page Temperature PRESSURE/STRAINFlow