Omega Engineering CYD201 and CYD208 manual Power and Ground Requirements

Page 14

OMEGA Model CYD201/CYD208 User’s Manual

2.3POWER AND GROUND REQUIREMENTS

The Model CYD201/CYD208 requires a power source of 90 to 125 or 210 to 250 VAC, 50 or 60 Hz, single phase, 3 Watts. Three-prong detachable power cord for 120 VAC operation included. Connect to rear panel UL/IEC/ICEE standard plug. See Table 2-1 for fuse rating.

WARNING: To prevent electrical fire or shock hazards, do not expose this instrument to rain or excess moisture.

CAUTION: Verify proper fuse installation and AC Line Voltage Selection Wheel on the Model CYD201/CYD208 rear panel set to available AC line voltage before inserting power cord and turning on the instrument.

Refer to Chapter 5 to change voltage configuration.

NOTE: Do not attach the shield to earth ground at the sensor end. It may introduce noise at the measurement end.

Table 2-1. Line Voltage and Fuse Rating Selection

Select

Range

Fuse

 

 

 

115

90 – 125 VAC

0.2 A (Slow Blow)

230

210 – 250 VAC

0.1 A (Slow Blow)

 

 

 

To protect operating personnel, the National Electrical Manufacturer’s Association (NEMA) recommends, and some local codes require, grounded instrument panels and cabinets. The 3-conductor power cable, when plugged into an appropriate receptacle, grounds the instrument.

Grounding and shielding signal lines are major concerns when setting up any precision instrument or system. The CYD201/CYD208 includes ground isolation of sensor excitation to allow 4-wire measurement of diode voltage and resistance. Improperly grounding sensor leads and shields can defeat this feature.

Model CYD201/CYD208 digital logic ties directly to earth ground for interface communication. The low side of the heater output connects directly to earth ground. Shield sensor cables whenever possible. Attach the shields to the connector shield pin.

2.4SENSOR INSTALLATION RECOMMENDATIONS

See the OMEGA Product Catalog for sensor installation and specifications. Call OMEGA for copies of application notes or sensor installation questions. Below are general recommendations on sensor installation:

1.Do not ground the sensor.

2.Shield leads and connect shield wire to SHIELD on screw terminal connector only. Do not connect shield at other end of cable.

2-2

Installation

Image 14
Contents M789-038A June Warranty Return Requests / InquiriesTable of Contents Table of Contents List of Illustrations This Page Intentionally Left Blank Table of Contents Model CYD201/CYD208 General Description Chapter IntroductionGeneral Model CYD201/CYD208 Specifications ResolutionHandling Liquid Helium and Liquid Nitrogen LHe and LN2 Safety PrecautionsHandling Cryogenic Storage Dewars Recommended First Aid Electrostatic DischargeHandling Esds Components Safety SummaryIdentifying Esds Components Ground The InstrumentKeep Away From Live Circuits Safety SymbolsDo Not Operate In An Explosive Atmosphere Do Not Substitute Parts Or Modify InstrumentRepackaging for Shipment Chapter InstallationInspection and Unpacking Power and Ground Requirements Sensor Installation RecommendationsFour-Lead Measurement Two-Lead Vs Four-Lead MeasurementsTwo-Lead Measurement Connecting Leads To The Sensor Sensor MountingMeasurement Errors Due To AC Noise Terminal Description Sensor Input ConnectionsJ1 Input Sensor Curve Definition Rack MountingHole Sizes Model CYD208-DIN Rack MountingInitial Power UP Sequence Power UP ErrorsThis Page Intentionally Left Blank Installation Chapter Operation Units KEYSetting Dwell Times Model CYD208 Only Alarm SetpointScan Mode Model CYD208 Only Alarm OperationSoftcal Compensations Latched And Unlatched AlarmsAlarm Fix Function Model CYD208 Only SoftCal Calibration Procedure Verifying SoftCal Operation Erasing SoftCal CompensationsOperation Serial Interface Chapter Remote OperationSerial Interface Connections Serial I/O RJ-11 Connector Pin Definitions J2 Serial I/OSerial Interface Operation Sample Basic ProgramInput Serial Interface Command SummarySample Quick Basic 4.0 Program ReturnedInput Lxxx.x Returned Nothing Input F0x Returned NothingInput Hxxx.x Returned Nothing Input R Returned NothingInput YCx Returned Nothing Switch ID and Alarm DataHigh or Low Parameter H = high alarm, L = low alarm End Scanning Model CYD208 OnlyScan Status Model CYD208 Only Sign,sensor reading,units,alarm status CrlfCurrent channel,sign,sensor reading,units, alarm statusCRLF This Page Intentionally Left Blank Remote Operation Chapter Service Model CYD201 Rear Panel ConnectionsModel CYD208 Rear Panel Connections Model CYD208 Rear Panel ConnectionsError Code Troubleshooting General MaintenanceFuse Replacement Line Voltage ConfigurationModel CYD201 Model CYD208 Current Source CalibrationRecalibration 2 A/D Converter Calibration Serial Interface Cable and AdaptersModel CYD200-B RJ-11 to DE-9 Adapter Wiring Details Page Chapter Options and Accessories Accessories Model Description of AccessoryDIN Model Description of AccessoryModel CYD201/CYD208 Wires Lsci P/N Description of Cable Series DT-450 Model CYD201/CYD208 Sensors Sensor no Description of SensorSeries DT-420 Series CY-7Options and Accessories Appendix a Curve Tables A1.0 GeneralTable A-2. Curve 1 DT-500DI-8A Voltage-Temp. Characteristics Table A-3. Curve 2 DT-500DRC-D Voltage-Temp. Characteristics 6591 18.0 Table A-5. Curve 4 CTI Diode Voltage-Temp. Characteristics Table A-6. Curve 5 DT-500DI-8C Voltage-Temp. Characteristics Table A-7. Curve 6 CY-7 Voltage-Temp. Characteristics 305.0 50689 Page Page Flow TemperaturePRESSURE/STRAIN