To within ±0.1K from 177 K to 313 K. To ±0.2 K or better from 30 K to 373 K. ±1.0 K above 373 K.
1.4 K to 475 K without probe
23 K to 473 K with probe
Connections: Four-lead sensor connection (2 current, 2 voltage).
Response Curves*: Standard Curve 10, DT-500DI-8A (also -8B and -8C), DT-500DRC-D, DT-500DRC-E1, and CTI Curve C.
* SOFTCAL™ qualified only for CY-7 Series diode sensors.
COMPUTER INTERFACE
Type: RS-232C Serial Three Wire (Refer to Table 4-1).
MECHANICAL
Ambient Temperature Range: 18 to 28 °C (64 to 82 °F), or 15 to 35 °C (59 to 95 °F) with reduced accuracy.
Power Requirements: 90-125 or 210-250 VAC, 50/60 Hz, 3 watts.
Dimensions: 41 x 106 x 164 mm (1.61 x 4.18 x 6.45 inches).
Weight: 0.5 kilogram (1.1 Pounds)
NOTES
1. Product Specifications subject to change without notice.
2. System electronic temperature accuracy in a given temperature range is the sum of the specifications given for input and output. Sensor calibration errors are not included.
Scan/Dwell:
Sensor Excitation: 10 µA constant current
Repeatability: <50 mK
Input Range: 0 to 3 volts with a resolution of 0.1 mV
Hi/Lo Alarm Setpoint: 0.1° resolution
Alarm Relay: Single SPDT relay, rated 28 VDC or Peak AC, 0.25 A (3 W max.)
System Accuracy:
Temperature Range:
0.1 for values > 100 or < –100
0.01 for values between -100<T<100
The Model CYD208 automatically scans all eight channels with selectable dwell times of 0 (skip), 5, 10, 30 and 60 seconds for each channel.
OMEGA Model CYD201/CYD208 User’s Manual
Table 1-1. Model CYD201/CYD208 Specifications
Display:
Resolution:
Introduction |