Omron ne-c801 instruction manual Using the Device

Page 17

USING THE DEVICE

5Turn the power switch to the off ( ) position.

The compressor turns off and nebulization stops.

6Disconnect the air tube from the nebulizer kit.

Hold the air tube plug and gently pull down to disconnect the air tube plug from the air tube connector on the bottom of the nebulizer kit.

7Check the air tube. No condensation or moisture should remain in the air tube.

WARNING

Do not store the air tube with moisture or medication remaining in the air tube. This could result in infection as a result of bacteria.

If any condensation or moisture remains in the air tube, remove the moisture from the air tube. Follow the directions below:

1.Make sure the air tube is still connected to the air tube connector on the upper side of the compressor.

2.Press the power switch on. The compressor will start and pump air through the air tube to expel the moisture.

3.Press the power switch again to turn the compressor off.

8Disconnect the air tube from the compressor. Hold the air tube plug and gently pull the air tube plug off the air tube connector on the upper side of the compressor.

9Unplug the AC adapter from the

electrical outlet.

CAUTION

• Unplug the AC adapter from the electrical outlet after using the device.

• Remove the AC adapter from the device after use.

Do not pull the power cord of AC adapter strongly.

17

Image 17
Contents NE-C801 Contents Staple your purchase receipt here IntroductionIntended Use Operating the Device Safety InformationSafety Information Risk of Electrical Shock Maintenance and Storage Nebulizer Kit Compressor Main Unit Know Your UnitComponents Know Your UnitPreparing the Nebulizer for Use Preparing the Nebulizer for USE Using the Adult Mask Optional Using the Nosepiece OptionalUsing the Child Mask Optional Attaching the Air Tube Hold the nebulizer kit as illustrated on the right Using the DeviceUsing the Child Mask or the Adult Mask Optional Using the DeviceUsing the Device Cleaning after Each Use Handling the Baffle Daily DisinfectingInsert a new air filter Changing the Air FilterCaring for the Device Cause Troubleshooting GuideLimited Warranty FCC Statement Contents SpecificationsParticle Specifications Technical DataElectromagnetic emissions IEC60601-1-2 Guidance and Manufacturers DeclarationElectromagnetic immunity IEC60601-1-2 Recommend separation distance KHz to 80 MHz KHz to 800 MHz MHz to 2.5GHz Page Omron Healthcare, Inc