Troubleshooting

Troubleshooting Guide

PROBLEM

Cause

SOLUTION

 

 

 

 

The AC adapter is not

Turn the power switch off.

No power to unit when

Plug the AC adapter into an

plugged into an electrical

the power switch is on.

electrical outlet. Turn the device

outlet.

 

on.

 

 

 

 

 

 

No medication in the

 

 

medication cup.

Add the correct amount of

 

Too much or too little

prescribed medication to the

 

medication in the

medication cup.

 

medication cup.

 

 

 

 

 

The baffle is not attached

Make sure the baffle is correctly

 

to the inhalation top or

 

attached to the inhalation top.

 

incorrectly positioned.

 

 

 

 

 

 

The nebulizer kit is not

Make sure the nebulizer kit is

 

correctly assembled and the

 

correctly assembled.

inhalation accessory is correctly

 

 

attached.

 

 

 

 

The nozzle is blocked.

Clean and disinfect the nebulizer kit

No nebulization or low

to remove the blockage.

 

nebulization rate when

The nebulizer kit is tilted at

Hold the nebulizer kit correctly.

the power is on.

Do not tilt the nebulizer kit so the

 

an incorrect angle.

angle of the kit is greater than 45

 

 

degrees.

 

 

 

 

The air tube is incorrectly

Make sure the air tube is correctly

 

attached to the compressor and the

 

attached.

 

nebulizer kit.

 

 

 

 

 

 

The air tube is folded

Make sure the air tube is not

 

folded, kinked or bent. Inspect the

 

or damaged.

 

air tube for any damage. Replace

 

The air tube is blocked.

 

the air tube if damaged.

 

 

 

 

 

 

The air filter is dirty.

Replace the air filter with a new

 

clean air filter.

 

 

 

 

 

The device is abnormally

The air filter cover is

Attach the air filter cover correctly.

Make sure the air filter cover is

loud.

incorrectly attached.

not blocked.

 

 

 

The compressor

Do not cover the compressor with

 

is covered.

any type of cover during use.

The device is very hot.

 

 

Operating continuously over

Limit use to 20 minutes at a time

 

 

and allow a 40 minutes interval

 

20 minutes.

 

before using the device again.

 

 

 

 

 

22

Image 22
Contents NE-C801 Contents Introduction Staple your purchase receipt hereIntended Use Safety Information Operating the DeviceSafety Information Risk of Electrical Shock Maintenance and Storage Know Your Unit Nebulizer Kit Compressor Main UnitKnow Your Unit ComponentsPreparing the Nebulizer for Use Preparing the Nebulizer for USE Using the Child Mask Optional Using the Nosepiece OptionalUsing the Adult Mask Optional Attaching the Air Tube Using the Device Hold the nebulizer kit as illustrated on the rightUsing the Device Using the Child Mask or the Adult Mask OptionalUsing the Device Cleaning after Each Use Daily Disinfecting Handling the BaffleChanging the Air Filter Insert a new air filterCaring for the Device Troubleshooting Guide CauseLimited Warranty FCC Statement Specifications ContentsTechnical Data Particle SpecificationsGuidance and Manufacturers Declaration Electromagnetic emissions IEC60601-1-2Electromagnetic immunity IEC60601-1-2 Recommend separation distance KHz to 80 MHz KHz to 800 MHz MHz to 2.5GHz Page Omron Healthcare, Inc