Troubleshooting
Troubleshooting Guide
PROBLEM | Cause | SOLUTION | |
|
|
| |
| The AC adapter is not | Turn the power switch off. | |
No power to unit when | Plug the AC adapter into an | ||
plugged into an electrical | |||
the power switch is on. | electrical outlet. Turn the device | ||
outlet. | |||
| on. | ||
|
| ||
|
|
| |
| No medication in the |
| |
| medication cup. | Add the correct amount of | |
| Too much or too little | prescribed medication to the | |
| medication in the | medication cup. | |
| medication cup. |
| |
|
|
| |
| The baffle is not attached | Make sure the baffle is correctly | |
| to the inhalation top or | ||
| attached to the inhalation top. | ||
| incorrectly positioned. | ||
|
| ||
|
|
| |
| The nebulizer kit is not | Make sure the nebulizer kit is | |
| correctly assembled and the | ||
| correctly assembled. | inhalation accessory is correctly | |
|
| attached. | |
|
|
| |
| The nozzle is blocked. | Clean and disinfect the nebulizer kit | |
No nebulization or low | to remove the blockage. | ||
| |||
nebulization rate when | The nebulizer kit is tilted at | Hold the nebulizer kit correctly. | |
the power is on. | Do not tilt the nebulizer kit so the | ||
| an incorrect angle. | angle of the kit is greater than 45 | |
|
| degrees. | |
|
|
| |
| The air tube is incorrectly | Make sure the air tube is correctly | |
| attached to the compressor and the | ||
| attached. | ||
| nebulizer kit. | ||
|
| ||
|
|
| |
| The air tube is folded | Make sure the air tube is not | |
| folded, kinked or bent. Inspect the | ||
| or damaged. | ||
| air tube for any damage. Replace | ||
| The air tube is blocked. | ||
| the air tube if damaged. | ||
|
| ||
|
|
| |
| The air filter is dirty. | Replace the air filter with a new | |
| clean air filter. | ||
|
| ||
|
|
| |
The device is abnormally | The air filter cover is | Attach the air filter cover correctly. | |
Make sure the air filter cover is | |||
loud. | incorrectly attached. | ||
not blocked. | |||
|
| ||
| The compressor | Do not cover the compressor with | |
| is covered. | any type of cover during use. | |
The device is very hot. |
|
| |
Operating continuously over | Limit use to 20 minutes at a time | ||
| |||
| and allow a 40 minutes interval | ||
| 20 minutes. | ||
| before using the device again. | ||
|
| ||
|
|
|