Omron ne-c801 instruction manual Recommend separation distance

Page 29

Guidance and Manufacturer's Declaration

Electromagnetic immunity IEC60601-1-2

Immunity test

IEC 60601

Compliance

Electromagnetic environment –

Test level

level

guidance

 

 

 

 

Portable and mobile RF

 

 

 

communications equipment should

 

 

 

be used no closer to any part of

 

 

 

the NE-C801 including cables,

 

 

 

than the recommended separation

 

 

 

distance calculated from the equation

 

 

 

appropriate to the frequency of the

 

 

 

transmitter.

 

 

 

Recommend separation distance

 

 

 

d = 1.2 √P

Conducted RF

3 V rms

3 V rms

d = 1.2 √P 80 MHz to 800 MHz

d = 2.3 √P 800 MHz to 2.5 GHz

IEC 61000-4-6

150 kHz ~

 

 

where P is the maximum output

 

80 MHz

 

 

 

power rating of the transmitter

 

 

 

Radiated RF

3 V/m

3 V/m

in watts (W) according to the

transmitter manufacturer and d is the

IEC 61000-4-3

80 MHz to

 

 

recommended separation distance in

 

2.5 GHz

 

 

 

meters (m).

 

 

 

 

 

 

Field strengths from fixed RF

 

 

 

transmitters as determined by an

 

 

 

electromagnetic site survey,*2) should

 

 

 

be less than the compliance level in

 

 

 

each frequency range.*3)

 

 

 

Interference may occur in the

 

 

 

vicinity of equipment marked with he

 

 

 

following symbol:

 

 

 

 

Note1: At 80 MHz and 800 MHz, the higher frequency range applies.

Note2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.

*2) Field strengths from fixed transmitters, such as base stations for radio (cellular/ cordless) telephones and land mobile radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the NE-C801 is used exceeds the applicable RF compliance level above, the NE-C801 should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the NE-C801.

*3) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

29

Image 29
Contents NE-C801 Contents Staple your purchase receipt here IntroductionIntended Use Operating the Device Safety InformationSafety Information Risk of Electrical Shock Maintenance and Storage Nebulizer Kit Compressor Main Unit Know Your UnitComponents Know Your UnitPreparing the Nebulizer for Use Preparing the Nebulizer for USE Using the Adult Mask Optional Using the Nosepiece OptionalUsing the Child Mask Optional Attaching the Air Tube Hold the nebulizer kit as illustrated on the right Using the DeviceUsing the Child Mask or the Adult Mask Optional Using the DeviceUsing the Device Cleaning after Each Use Handling the Baffle Daily DisinfectingInsert a new air filter Changing the Air FilterCaring for the Device Cause Troubleshooting GuideLimited Warranty FCC Statement Contents Specifications Particle Specifications Technical DataElectromagnetic emissions IEC60601-1-2 Guidance and Manufacturers DeclarationElectromagnetic immunity IEC60601-1-2 Recommend separation distance KHz to 80 MHz KHz to 800 MHz MHz to 2.5GHz Page Omron Healthcare, Inc
Related manuals
Manual 30 pages 58.15 Kb Manual 30 pages 40.59 Kb