Guidance and Manufacturer's Declaration

Recommended separation distance between portable and mobile RF communications equipment and the NE-C801

The NE-C801 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customers or the users of the NE-C801 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the NE-C801 as recommended below, according to the maximum output power of the communications equipment.

Output Power of

Separation distance according to frequency of transmitter in meter

150 kHz to 80 MHz

80 kHz to 800 MHz

800 MHz to 2.5GHz

Transmitter in Watt

d = 1.2 √P

d = 1.2 √P

d = 2.3 √P

 

0.01

0.12

0.12

0.23

0.1

0.38

0.38

0.73

1

1.2

1.2

2.3

10

3.8

3.8

7.3

100

12

12

23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

Note: At 80MHz and 800MHz, the separation distance for the higher frequency range applies

Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.

EMC tests have included the AC adapter as included with the product.

30

Image 30
Contents NE-C801 Contents Introduction Staple your purchase receipt hereIntended Use Safety Information Operating the DeviceSafety Information Risk of Electrical Shock Maintenance and Storage Know Your Unit Nebulizer Kit Compressor Main UnitKnow Your Unit ComponentsPreparing the Nebulizer for Use Preparing the Nebulizer for USE Using the Nosepiece Optional Using the Child Mask OptionalUsing the Adult Mask Optional Attaching the Air Tube Using the Device Hold the nebulizer kit as illustrated on the rightUsing the Device Using the Child Mask or the Adult Mask OptionalUsing the Device Cleaning after Each Use Daily Disinfecting Handling the BaffleChanging the Air Filter Insert a new air filterCaring for the Device Troubleshooting Guide CauseLimited Warranty FCC Statement Specifications ContentsTechnical Data Particle SpecificationsGuidance and Manufacturers Declaration Electromagnetic emissions IEC60601-1-2Electromagnetic immunity IEC60601-1-2 Recommend separation distance KHz to 80 MHz KHz to 800 MHz MHz to 2.5GHz Page Omron Healthcare, Inc