Daily Disinfecting

Disinfect the nebulizer kit and mouthpiece, or optional masks or optional nosepiece after the last treatment of the day.

To disinfect the parts, use one of the methods described below:

A.Use a commercially available disinfectant. Follow the instructions provided by the disinfectant manufacturer.

1 Submerge the parts in the cleansing solution for the specified period.

2 Remove the parts and discard the solution.

3 Rinse the parts with clean hot tap water, shake off excess water and allow to air dry in a clean environment.

B.Parts may be boiled between 15 to 20 minutes.

After boiling, carefully remove the parts, shake off excess water and allow to air dry in a clean environment.

NOTE: Do not boil air tube, air filter, air filter cover and optional masks.

CAUTION

Do not use an autoclave, EOG gas disinfection or low temperature plasma sterilizer to disinfect the device.

Handling the Baffle

The baffle is an important part that is used to nebulize medication. Be sure to observe the following precautions when handling it.

CAUTION

Always wash the baffle after each use.

Do not use a brush or pin etc., to clean the parts.

• When disinfecting the parts by boiling, be sure to boil them in plenty of water.

Do not boil the baffle together with other objects, other than applicable nebulizer accessories.

Do not use a microwave oven, dish dryer or hair dryer to dry parts.

19

Image 19
Contents NE-C801 Contents Staple your purchase receipt here IntroductionIntended Use Operating the Device Safety InformationSafety Information Risk of Electrical Shock Maintenance and Storage Nebulizer Kit Compressor Main Unit Know Your UnitComponents Know Your UnitPreparing the Nebulizer for Use Preparing the Nebulizer for USE Using the Child Mask Optional Using the Nosepiece OptionalUsing the Adult Mask Optional Attaching the Air Tube Hold the nebulizer kit as illustrated on the right Using the DeviceUsing the Child Mask or the Adult Mask Optional Using the DeviceUsing the Device Cleaning after Each Use Handling the Baffle Daily DisinfectingInsert a new air filter Changing the Air FilterCaring for the Device Cause Troubleshooting GuideLimited Warranty FCC Statement Contents SpecificationsParticle Specifications Technical DataElectromagnetic emissions IEC60601-1-2 Guidance and Manufacturers DeclarationElectromagnetic immunity IEC60601-1-2 Recommend separation distance KHz to 80 MHz KHz to 800 MHz MHz to 2.5GHz Page Omron Healthcare, Inc