Guidance and Manufacturer's Declaration
Electromagnetic immunity IEC60601-1-2
Immunity test | IEC 60601 | Compliance | Electromagnetic environment – | |
Test level | level | guidance | ||
| ||||
Electrostatic | ±6 kV | ±6 kV | Floor should be wood, concrete, or | |
ceramic tile. If floors are covered | ||||
discharge (ESD) | contact | contact | ||
with synthetic material, the relative | ||||
IEC | ±8 kV air | ±8 kV air | ||
humidity should be at least 30 %. | ||||
|
|
| ||
| ±2 kV |
|
| |
Electrical fast | for power | ±2 kV | Mains power quality should be that of | |
supply lines | for power | |||
transient/burst | a typical commercial and/or hospital | |||
±1 kV for | supply lines | |||
IEC | environment. | |||
input/output | *1) | |||
|
| |||
| lines |
|
| |
| ±1 kV line | ±1 kV line | Mains power quality should be that of | |
Surge | to line | to line | ||
a typical commercial and/or hospital | ||||
IEC | ±2 kV line | ±2 kV line | ||
environment. | ||||
| to earth | to earth | ||
|
| |||
| <5 % UT | <5 % UT |
| |
| (>95 % dip | (>95 % dip |
| |
| in UT) for | in UT) for |
| |
| 0.5 cycle | 0.5 cycle | Mains power quality should be that of | |
| 40 % UT | 40 % UT | ||
| a typical commercial and/or hospital | |||
| (60 % dip | (60 % dip | ||
Voltage dips, short | environment. | |||
in UT) for 5 | in UT) for 5 | |||
interruptions and | If the user of the | |||
cycles | cycles | |||
voltage variations on | continued operation during power | |||
70 % UT | 70 % UT | |||
power supply | mains interruption, it is recommended | |||
(30 % dip in | (30 % dip in | |||
IEC | that the | |||
UT) for 25 | UT) for 25 | |||
| an uninterruptible power supply or | |||
| cycles | cycles | ||
| battery. | |||
| <5 % UT | <5 % UT | ||
|
| |||
| (95 % dip | (95 % dip |
| |
| in UT) for 5 | in UT) for 5 |
| |
| sec. | sec. |
| |
Power frequency |
|
| Power frequency magnetic fields | |
(50/ 60 Hz) magnetic | 3 A/m | 3 A/m | should be at levels characteristic | |
field | of a typical location in a typical | |||
|
| |||
IEC |
|
| commercial or hospital environment. |
Note: UT is the A.C. mains voltage prior to application of the test level.
*1) The test of input/output lines is not applicable since they are shorter than 3.0m.