Omron ne-c801 instruction manual Electromagnetic immunity IEC60601-1-2

Page 28

Guidance and Manufacturer's Declaration

Electromagnetic immunity IEC60601-1-2

Immunity test

IEC 60601

Compliance

Electromagnetic environment –

Test level

level

guidance

 

Electrostatic

±6 kV

±6 kV

Floor should be wood, concrete, or

ceramic tile. If floors are covered

discharge (ESD)

contact

contact

with synthetic material, the relative

IEC 61000-4-2

±8 kV air

±8 kV air

humidity should be at least 30 %.

 

 

 

 

±2 kV

 

 

Electrical fast

for power

±2 kV

Mains power quality should be that of

supply lines

for power

transient/burst

a typical commercial and/or hospital

±1 kV for

supply lines

IEC 61000-4-4

environment.

input/output

*1)

 

 

 

lines

 

 

 

±1 kV line

±1 kV line

Mains power quality should be that of

Surge

to line

to line

a typical commercial and/or hospital

IEC 61000-4-5

±2 kV line

±2 kV line

environment.

 

to earth

to earth

 

 

 

<5 % UT

<5 % UT

 

 

(>95 % dip

(>95 % dip

 

 

in UT) for

in UT) for

 

 

0.5 cycle

0.5 cycle

Mains power quality should be that of

 

40 % UT

40 % UT

 

a typical commercial and/or hospital

 

(60 % dip

(60 % dip

Voltage dips, short

environment.

in UT) for 5

in UT) for 5

interruptions and

If the user of the NE-C801 requires

cycles

cycles

voltage variations on

continued operation during power

70 % UT

70 % UT

power supply

mains interruption, it is recommended

(30 % dip in

(30 % dip in

IEC 61000-4-11

that the NE-C801 be powered from

UT) for 25

UT) for 25

 

an uninterruptible power supply or

 

cycles

cycles

 

battery.

 

<5 % UT

<5 % UT

 

 

 

(95 % dip

(95 % dip

 

 

in UT) for 5

in UT) for 5

 

 

sec.

sec.

 

Power frequency

 

 

Power frequency magnetic fields

(50/ 60 Hz) magnetic

3 A/m

3 A/m

should be at levels characteristic

field

of a typical location in a typical

 

 

IEC 61000-4-8

 

 

commercial or hospital environment.

Note: UT is the A.C. mains voltage prior to application of the test level.

*1) The test of input/output lines is not applicable since they are shorter than 3.0m.

28

Image 28
Contents NE-C801 Contents Introduction Staple your purchase receipt hereIntended Use Safety Information Operating the DeviceSafety Information Risk of Electrical Shock Maintenance and Storage Know Your Unit Nebulizer Kit Compressor Main UnitKnow Your Unit ComponentsPreparing the Nebulizer for Use Preparing the Nebulizer for USE Using the Child Mask Optional Using the Nosepiece OptionalUsing the Adult Mask Optional Attaching the Air Tube Using the Device Hold the nebulizer kit as illustrated on the rightUsing the Device Using the Child Mask or the Adult Mask OptionalUsing the Device Cleaning after Each Use Daily Disinfecting Handling the BaffleChanging the Air Filter Insert a new air filterCaring for the Device Troubleshooting Guide CauseLimited Warranty FCC Statement Specifications ContentsTechnical Data Particle SpecificationsGuidance and Manufacturers Declaration Electromagnetic emissions IEC60601-1-2Electromagnetic immunity IEC60601-1-2 Recommend separation distance KHz to 80 MHz KHz to 800 MHz MHz to 2.5GHz Page Omron Healthcare, Inc