Health O Meter 2650KL operation manual Calibration Path, Operator Action Display, Display Assembly

Page 21

CALIBRATION PATH

The calibration is performed using kg or lb, according to the units of measure used upon entering into the calibration path.

 

Operator Action

Display

1.

Press ON/OFF button and immediately press and

Calibration

 

hold HOLD/RELEASE button for 3-4 seconds

Process

2.

Press ENTER

Enter load weight

 

 

600.0

3.

Using cd keys and/or the keypad, set the calibration

Zero calibration

 

load to at least 200 lbs (for best results and greater

Please clear the scale

 

accuracy, use a 600 lb weight). Press ENTER

 

 

 

 

4.

Please clear the scale platform and press ENTER

Zero calibration

 

 

Please wait

5.

Please wait without touching or shaking the scale

Weight calibration

 

until the zero calibration process is finished

Put: XXX.X

6.

Load the scale platform with the required weight and

Weight calibration

 

press ENTER

Please wait

7.

Please wait without touching or shaking the scale

Calibration

 

until the calibration process is finished

factor: X.XXXXX

8.

Press ENTER

Calibration

 

 

Please clear the scale

9.

Please remove weight from the scale platform

Rebooting

 

 

Please wait

10.

Please wait until the scale resumes normal operation

Health O Meter Pro Plus

EXPLODED VIEW OF SCALE

Display Assembly

20

Image 21
Contents PRO Plustm Electronic Wheelchair Scale Table of Contents Specifications Scale SpecificationsGeneral Assembly after Shipment Assembly InstructionsRefer to the Parts List on page 22 for details on parts Folded Ramp SET UP Preparing the Scale for UseReplacing Batteries Operation Instructions Description Function Weight Display Mode MenuOnly applies to pound values, not the metric values Tare Weight Data ManagementRemoving the Tare Enter Scale Settings Operation Instructions System Settings System Test User Settings Operation Instructions Maintenance MaintenanceCleaning Disconnect the scale from the AC power sourceSymptom Possible Cause TroubleshootingDisplay Assembly Calibration PathOperator Action Display Exploded View of Scale Scale Platform and Ramps CAT No Parts ListLimited Warranty WarrantyWhat does the Warranty Cover? To get Warranty ServiceTable DES Matieres Precaution ET Prevention Specifications DE LA BalanceGeneralites Montage après envoi Mode DassemblageMode Dassemblage Pliage de la rampes daccès Installation Remplacement des pilesPréparation de la balance avant usage Mode DE Fonctionnement Touche Description Fonction Menu Principal Mode Daffichage DU PoidsPoids DE LA Tare Gestion DES DonneesRetrait de la Tare Mode DE Fonctionnement Reglages DE LA Balance Mode DE Fonctionnement Reglages DU Systeme Test DE Systeme Reglages Utilisateur Mode DE Fonctionnement Déconnectez la balance du secteur CA NettoyageLocalisation DES Pannes Symptome Cause Possible Action CorrectiveAction réalisée par lopérateur Message affiché CalibragePage Liste DES Pieces Garantie Pour obtenir un service de garantieÍndice Precaución Y Advertencia EspecificacionesEspecificaciones DE LA Balanza Figura 1. Montaje Después del Envío Instrucciones DE MontajeFigura Figura 3. Moviendo la Rampa Configuración Preparando la Balanza para ser UsadaCambiando las Baterías Instrucciones DE Operación Ítem Descripción Función Modo DE Presentación DEL Peso MenúPeso Tara Administración DE DatosTara Automática Removiendo la TaraCalculando EL BMI Configuraciones DE LA Balanza Instrucciones DE Operación Configuraciones DEL Sistema Prueba DEL Sistema Configuraciones DEL Usuario Instrucciones DE Operación Mantenimiento MantenimientoLimpieza Desconecte la balanza de la fuente de energía CASíntoma Posible Causa Acción Correctiva Solución DE ProblemasAcción del Operador Display Trayectoria DE CalibraciónPage Lista DE Partes Garantía Garantía Limitada¿Qué Cubre la Garantía? AA0367-0 August