SECTION 1—GENERAL GUIDELINES
SECTION 1—GENERAL GUIDELINES
WARNING
SECTION 1 - GENERAL GUIDELINES contains important information for the safe operation and use of this product.
Repair or Service Information
Set‐up of the Electronics Control Unit is to be performed only by a qualified technician. The final adjustments of the controller may affect other activities of the wheelchair. Damage to the equipment could occur if improperly set‐up or adjusted.
Except for programming, DO NOT service or adjust the wheelchair while occupied, unless otherwise noted.
A pinch point exists between walking beam/head tube cap and telescoping tube when TDX is at the lowest seat to floor height.
Before adjusting, repairing or servicing the wheelchair, ALWAYS turn the wheelchair power Off, otherwise, injury or damage may occur.
Invacare products are specifically designed and manufactured for use in conjunction with Invacare accessories. Accessories designed by other manufacturers have not been tested by Invacare and are not recommended for use with Invacare products.
Transport ready packages are not retrofittable to existing models and are not field serviceable.
Battery retention brackets MUST be installed at all times. Otherwise, the wheelchair will not be WC/19 compliant. Refer to Removing/Installing the Batteries From/Into the Wheelchair on page 55.
Wheelchairs should be examined during maintenance for signs of corrosion (water exposure, incontinence, etc.). Electrical components damaged by corrosion should be replaced IMMEDIATELY.
Wheelchairs that are used by incontinent users and/or are frequently exposed to water may require replacement of electrical components more frequently.
For optimum performance, replace gas‐locking cylinders every 2 years.
Operation Information
Performance adjustments should only be made by professionals of the healthcare field or persons fully conversant with this process and the driverʹs capabilities. Incorrect settings could cause injury to the driver, bystanders, damage to the wheelchair and to surrounding property.
After the wheelchair has been set‐up/adjusted, check to make sure that the wheelchair performs to the specifications entered during the set‐up procedure. If the wheelchair does NOT perform to specifications, turn the wheelchair Off immediately and reenter set‐up specifications. Repeat this procedure until the wheelchair performs to specifications.
TDX® SI, TDX | 12 | Part No 1154294 |