Omega DP26-RTD manual Front Panel Lockout Push Button Lockout, Jumper Lockout

Page 16

2.5 FRONT PANEL LOCKOUT

2.5.1 PUSH BUTTON LOCKOUT

To lock the RESET, MENU and SETPTS buttons, verify that the S3-A jumper is installed and the S3-E jumper is not installed, then follow these steps:

In the run mode-

1.Press and hold down the RESET button. The meter shows “SP.RS”. Do not release the RESET button.

2.Press the MENU button and hold RESET button down (3-5 seconds) until the meter shows “LOCK”.

3.Release both buttons.

To unlock the RESET, MENU and SETPTS buttons, follow these steps:

1.Press and hold down the RESET button. The meter shows “LOCK”. Do not release the RESET button.

2.Press the MENU button and hold RESET button down (3-5 seconds) until the meter shows “UN.LK”.

3.Release both buttons.

2.5.2 JUMPER LOCKOUT

To lock all front panel buttons, remove the S3-A jumper (refer to Figure 3-2).

To lock the MENU button only, verify that the S3-B jumper is removed, then install the S3-A and S3-E jumpers, (refer to Table 3-1).

8

Image 16
Contents User’s Guide Servicing North America Standard Procedures PrefacePage Iii Table of Contents Selecting Output Configuration OT.CF List of Figures Introduction UnpackingSafety Considerations EMC ConsiderationsAbout the Meter Standard FeaturesDescription Available Accessories Add-On OptionsFront of the Meter Shows the location of each part of the front of the meterMeter Buttons Setpts ButtonReset Button Front Panel Lockout Push Button Lockout Jumper LockoutConnector label ac Powered and dc Powered detail Back of the MeterBack of the Meter Connector Description Disassembly Disconnect the power supply before proceedingGetting Started Main Board Power JumpersRATING/PRODUCT Label Main Board Jumper Positions S3 Jumper Functions Mounting the Meter PlcsConnecting Sensor Input Through 3-7 illustrate how to connect you power sensorsConnecting Main Power Main Power Connections ac powered unitMain Power Connections dc powered unit Analog and Relay Output Connections11 Analog Output Connections Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Menu /DEV Reset DescriptionDisplay /DEV Setpoint Configuration Displays SetpointSpecifications Signal InputInput Power Information Analog Output if applicable Side View TOP View Factory Preset Values Menu Item Factory Preset ValuesCE Approvals Information WARRANTY/DISCLAIMER OMEGA…Of Course