3.7 Main Power Connections

Connect the ac main power connections as shown below.

WARNING: Do not connect ac power to your meter until you have completed all input and output connections. Failure to do so may result in injury! This device must only be installed electrically by specially trained electrician with corresponding qualifications. The main power input to the unit as well as the AC input signal to be measured must agree with the wiring instruction.

The meter is factory set to the power specified by the customer at the time of ordering. The voltage is printed on the Label under Input power.

Refer to Section 3-2to convert from one Note configuration to another

To connect power to the unit, attach wires from the ~AC power cord to the orange 3-pin female connector that is supplied with the meter. Refer to Figure 3-11for wiring instructions. Plug the female connector into the orange P1 (male) connector at the rear of the meter.

Figure 2Ð2 shows the location of P1 on the meter.

EARTH

NEUTRAL

LINE

GREEN WIRE ©

N

LP1

FUSE

SWITCH

Check for proper Earth grounding in the power distribution system (single phase).

1

ORANGE

2

3

CONNECTOR

Screws

underneath

Figure 3-11. ~AC Connector Wiring (P1)

16

Image 23
Contents DP119-RTD For immediate technical or application assistance Table of Contents Section Preface Model Type ResolutionModelpower Models Available RTD Models AvailableOther Models Available Page QTY Description Introduction UnpackingUnpacking & Inspection Safety ConsiderationsFront of the Meter About the Meter+EXC PIN #EXC Getting Started 3.1 Main Board Power Jumpers 1shows the location of solder jumpers W1 through W3 AC Voltage Install Remove2shows the location of the jumper plugs S1 through S5 Changing the Meter ParametersFollowing tables summarizes what parameters you can change Sensor Break Protection Example of How to Set Meter Option MF1 Jumpers S4 and S5Installation and Panel Mounting Installation Information Sensor Input Connections Wire RTDWire colors are for Omega RTD wiring only Analog Output Connections P2 Analog Output ConnectionsDisplay Hold Connections 2shows the location of the P2 connector on the meterÐ2 shows the location of P1 on the meter Main Power Connections2shows the location of P1 on the meter DC WiringDisassembly/Assembly Pull TAB Here PC Board Push Here Equipment Required Operation and CalibrationWire Configuration Calibration ProcedureZero Specifications Display Dimensions Glossary Page USAWARRANTY/DISCLAIMER Temperature