Omega DP119-RTD manual Display

Page 31

DISPLAY

Type:

7-segment, LED

Height:

0.56" (14.2 mm)

Symbols:

-1.8.8.8

Overrange Indication:

Three least-significant digits blank

Colors:

Red - standard

 

 

 

Green - optional

 

 

 

 

DIGITAL INPUTS

 

 

 

 

 

Hold:......................................................................TTL or 5V CMOS compatible

ENVIRONMENTAL

Operating Temperature:

...............................................0¡ to 60¡C(32¡ to 140¡F)

Storage Temperature:

-40¡ to 85¡C (-40¡ to 185¡F)

Relative Humidity:

95% at 40¡C (non-condensing)

MECHANICAL DIMENSIONS

Bezel:

01.89Ó x 3.78" (48 x 96 mm)

Depth Behind the Bezel:

4.32" (110 mm) w/ connectors

Panel Cutout:

1/8 DIN: 01.72" H x 3.62" W (45 x 92 mm)

Weight:

8 ounces (227 grams)

Panel Thickness

minimum: 0.03" (0.76mm)

 

maximum: 0.25" (6.25mm)

Case Material:

94V-1 UL-rated thermoplastic

POWER

AC Frequency:

50/60 Hz

Max. Power:

2.4 watts max.

AC Voltage:

115Vac ±15% 25mA max.

 

230Vac ±15% 14mA max.

 

100Vac ±15% 30mA max.

 

24Vac ±15% 127mA max.

DC Voltage:

9-26 Vdc @ 110mA

External Fuse Protection needed:

32mA, 115Vac/230Vac

UL Slow-blow (per single unit)

40mA, 100Vac

 

160mA, 24Vac

 

125mA, 9-26Vdc

Installation Category:

I per EN61010

Equipment Class:

I per EN61010

Pollution Degree:

2 per EN61010

Refer to the Glossary in Section 6.

 

24

Image 31
Contents DP119-RTD For immediate technical or application assistance Table of Contents Section Preface Model Type ResolutionModelpower Models Available RTD Models AvailableOther Models Available Page QTY Description Introduction UnpackingUnpacking & Inspection Safety ConsiderationsFront of the Meter About the MeterEXC PIN #+EXC Getting Started 3.1 Main Board Power Jumpers 1shows the location of solder jumpers W1 through W3 AC Voltage Install Remove2shows the location of the jumper plugs S1 through S5 Changing the Meter ParametersFollowing tables summarizes what parameters you can change Sensor Break Protection Example of How to Set Meter Option MF1 Jumpers S4 and S5Installation and Panel Mounting Installation Information Sensor Input Connections Wire RTDWire colors are for Omega RTD wiring only Analog Output Connections P2 Analog Output ConnectionsDisplay Hold Connections 2shows the location of the P2 connector on the meterÐ2 shows the location of P1 on the meter Main Power Connections2shows the location of P1 on the meter DC WiringDisassembly/Assembly Pull TAB Here PC Board Push Here Equipment Required Operation and CalibrationWire Configuration Calibration ProcedureZero Specifications Display Dimensions Glossary Page USAWARRANTY/DISCLAIMER Temperature