4.2CALIBRATION PROCEDURE
1.Remove the front lens. Connect power to the meter, observing proper wiring procedures.
2.Connect the precision resistance source to the P3 connector observing proper wiring configuration. For 3Ðwire connections, use pins 1, 2, 3, and 4. Refer to Figures
3.Apply resistance input as indicated in Table
4.Adjust either the ZERO or SPAN potentiometer as shown in Figure
5.After adjusting the post, go back and verify both temperature readings. Repeat step 4 as required.
Note ☞ Refer to Table
P3
+ Excitation | 4 |
Signal Hi | 3 |
REDRED
PRECISION RESISTANCE
Signal Lo | 2 |
- Excitation |
|
BLACK
BLACK
SOURCE
1 |
JUMPER WIRE (24 AWG)
Figure 4-1. 2-Wire Configuration
P3 |
| NOT USED |
| ||
+ Excitation | 4 |
| |||
RED |
|
|
| ||
|
|
| |||
Signal Hi | 3 |
|
|
| |
BLACK |
| ||||
Signal Lo | 2 |
| |||
- Excitation | 1 | BLACK |
|
| |
|
RED PRECISION RESISTANCE
BLACK SOURCE
Figure 4-2. 3-Wire Configuration
21